Post-processing image filtration enabling dose reduction in standard abdominal CT
Lund University.
Malmö University Hospital.
Malmö University Hospital.
Linköpings universitet, Institutionen för klinisk och experimentell medicin. Linköpings universitet, Hälsouniversitetet.
2010 (Engelska)Ingår i: RADIATION PROTECTION DOSIMETRY, ISSN 0144-8420, Vol. 139, nr 1-3, s. 180-185Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

The collective effective radiation dose to the population is increasing due to a higher use of computerised tomography. SharpView AB, Linkoping, Sweden, has developed an adaptive non-linear post-processing image filtration that may enable the use of lower radiation doses. The present study assessed if a lower dose with image filtration had the same image quality as a higher dose without the filter applied. All imaging was performed on a Siemens Somatom Sensation 16 CT. The parameters used were 120 kV and 200 mAs (40 patients) and 130 mAs without and with image filtering (40 patients), respectively. All studies were quantitatively evaluated for noise and image quality was assessed by visual grading characteristics (VGC) analysis. After image filtration, the noise in the processed images was lowered and the image quality was improved as shown by the VGC analysis. However, images using the higher dose were still ranked as the best in five out of eight criteria as shown by the VGC analysis. Image filtration enhances CT images significantly and further studies will show if 130 mAs with image filtration may be sufficient for clinically general abdominal CT.

Ort, förlag, år, upplaga, sidor
Oxford University Press , 2010. Vol. 139, nr 1-3, s. 180-185
Nationell ämneskategori
Teknik och teknologier
Identifikatorer
URN: urn:nbn:se:liu:diva-57154DOI: 10.1093/rpd/ncq086ISI: 000277738200031OAI: oai:DiVA.org:liu-57154DiVA, id: diva2:323571
Tillgänglig från: 2010-06-11 Skapad: 2010-06-11 Senast uppdaterad: 2010-06-11

Open Access i DiVA

Fulltext saknas i DiVA

Övriga länkar

Förlagets fulltext

Personposter BETA

Albertsson, Ida

Sök vidare i DiVA

Av författaren/redaktören
Albertsson, Ida
Av organisationen
Institutionen för klinisk och experimentell medicinHälsouniversitetet
Teknik och teknologier

Sök vidare utanför DiVA

GoogleGoogle Scholar

doi
urn-nbn

Altmetricpoäng

doi
urn-nbn
Totalt: 25 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf