To clinically measure blood pressure at extra corporeal bloodlines involves a hazard due to the infection risk and a risk for thrombosis formation. The aim was to design a non-invasive pressure sensor, measuring directly on a tube section. A modified tube cross-section was used to improve sensitivity. Using the developed sensing principle, a consistent relation (r=0.999) was obtained between pressure and output signal. The output was stable and an acceptable drift within the temperature-range. The method shows great promise for applications in monitoring of the dialysis process.
23rd Annual International Conference of the IEEE-Engineering-in-Medicine-and-Biology-Society, ISTANBUL, TURKEY, OCT 25-28, 2001