INTRODUCTION
The success of the deep brain stimulation (DBS) therapy relies primarily in the localization of the implanted electrode, implying the need of utmost accuracy in the targeting process. Intraoperative microelectrode recording and stimulation tests are a common procedure before implanting the permanent DBS lead to determine the optimal position with a large therapeutic window where side effects are avoided and the best improvement of the symptoms is achieved. Differences in dimensions and operating modes exist between the exploration and the permanent DBS electrode which might lead to different stimulation fields, even when ideal placement is achieved. The aim of this investigation is to compare the electric field (EF) distribution around the intraoperative and the chronic electrode, assuming that both have exactly the same position.
METHODS
3D models of the intraoperative exploration electrode and the chronically implanted DBS lead 3389 (Medtronic Inc., USA) were developed using COMSOL 5.2 (COMSOL AB, Sweden). Patient-specific MR images were used to determine the conductive medium around the electrode. The exploration electrode and the first DBS contact were set to current and voltage respectively (0.2mA(V) - 3 mA(V) in 0.1 mA(V) steps). The intraoperative model included the grounded guide tube used to introduce the exploration electrode; for the chronic DBS model, the outer boundaries were grounded and the inactive contacts were set to floating potential considering a monopolar configuration. The localization of the exploration and the chronic electrode was set according to the planned trajectory. The EF was visualized and compared in terms of volume and extension using a fixed isocontour of 0.2 V/mm.
RESULTS
The EF distribution simulated for the exploration electrode showed the influence of the parallel trajectory and the grounded guide tube. For an amplitude of e.g. 2 mA/2 V, the EF extension of the intraoperative was 0.6 mm larger than the chronic electrode at the target level; the corresponding difference in volume was 76.1 mm3.
CONCLUSION
Differences in the EF shape between the exploration and the chronic DBS electrode have been observed using patient-specific models. The larger EF extension obtained for the exploration electrode responds to its higher impedance and the use of current controlled stimulation. The presence of EF around the guide tube and the influence of the parallel trajectory require further experimental and clinical evaluation.