Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction
Linköping University, Center for Medical Image Science and Visualization (CMIV). Linköping University, Department of Medical and Health Sciences, Division of Radiological Sciences. Region Östergötland, Center for Diagnostics, Department of Radiology in Linköping. Linköping University, Faculty of Medicine and Health Sciences.ORCID iD: 0000-0002-4013-4615
Linköping University, Department of Medical and Health Sciences, Division of Radiological Sciences. Linköping University, Faculty of Medicine and Health Sciences. Region Östergötland, Center for Diagnostics, Medical radiation physics.
School of Technology and Health (STH), KTH Royal Institute, Stockholm, Sweden.ORCID iD: 0000-0002-7750-1917
Linköping University, Department of Medical and Health Sciences, Division of Radiological Sciences. Linköping University, Faculty of Medicine and Health Sciences. Linköping University, Center for Medical Image Science and Visualization (CMIV). Region Östergötland, Center for Diagnostics, Department of Radiology in Linköping.ORCID iD: 0000-0002-9446-6981
Show others and affiliations
2018 (English)In: European Radiology, ISSN 0938-7994, E-ISSN 1432-1084Article in journal (Refereed) Published
Abstract [en]

Purpose To estimate potential dose reduction in abdominal CT by visually comparing images reconstructed with filtered back projection (FBP) and strengths of 3 and 5 of a specific MBIR.

Material and methods A dual-source scanner was used to obtain three data sets each for 50 recruited patients with 30, 70 and 100% tube loads (mean CTDIvol 1.9, 3.4 and 6.2 mGy). Six image criteria were assessed independently by five radiologists. Potential dose reduction was estimated with Visual Grading Regression (VGR).

Results Comparing 30 and 70% tube load, improved image quality was observed as a significant strong effect of log tube load and reconstruction method with potential dose reduction relative to FBP of 22–47% for MBIR strength 3 (p < 0.001). For MBIR strength 5 no dose reduction was possible for image criteria 1 (liver parenchyma), but dose reduction between 34 and 74% was achieved for other criteria. Interobserver reliability showed agreement of 71–76% (κw 0.201–0.286) and intra-observer reliability of 82–96% (κw 0.525–0.783).

Conclusion MBIR showed improved image quality compared to FBP with positive correlation between MBIR strength and increasing potential dose reduction for all but one image criterion.

Place, publisher, year, edition, pages
Heidelberg: Springer, 2018.
Keywords [en]
Dose Computed tomography Iterative reconstruction Abdomen FBP
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:liu:diva-145274DOI: 10.1007/s00330-017-5113-4ISI: 000431653200023PubMedID: 29368163Scopus ID: 2-s2.0-85040915759OAI: oai:DiVA.org:liu-145274DiVA, id: diva2:1183967
Note

Funding agencies: ALF-grant from Region Ostergotland; LFoU-grant from Region Ostergotland; Medical Faculty at Linkoping University

Available from: 2018-02-20 Created: 2018-02-20 Last updated: 2019-10-15Bibliographically approved
In thesis
1. Visual grading evaluation of reconstruction methods and dose optimisation in abdominal Computed Tomography
Open this publication in new window or tab >>Visual grading evaluation of reconstruction methods and dose optimisation in abdominal Computed Tomography
2019 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Since its introduction in the 1970’s CT has emerged as a modality of choice because of its high sensitivity in producing accurate diagnostic images. A third of all Computed Tomography (CT) examinations are abdominal CTs which deliver one of the highest doses among common examinations. An increase in the number of CT examinations has raised concerns about the negative effects of ionising radiation as the dose is cumulative over the life span of the individual. Image quality in CT is closely related to the radiation dose, so that a certain dose with an associated small, but not negligible, risk is a prerequisite for high image quality. Typically, dose reduction in CT results in higher noise and a decrease in low contrast resolution which can be detrimental to the image quality produced. New technology presents a wide range of dose reduction strategies, the latest being iterative reconstruction (IR).The aim of this thesis was to evaluate two different classes of iterative reconstruction algorithms: statistical (SAFIRE) and model-based (ADMIRE) as well as to explore the diagnostic value of a low-dose abdominal CT for optimisation purposes.

This thesis included a total of 140 human subjects in four image quality evaluation studies, three of which were prospective studies (Papers I, II and IV) and one retrospective study (Paper III). Visual grading experiments to determine the potential dose reductions, were performed with pairwise comparison of image quality in the same patient at different tube loads (dose) and reconstructed with Filtered back projection (FBP) and SAFIRE strength 1 in a low-dose abdominal CT (Paper I) and FBP and ADMIRE strengths 3 and 5 in a standard dose abdominal CT (Paper II). Paper IV evaluated the impact of slice thicknesses in CT images reconstructed with ADMIRE strengths 3 and 5 when comparing multiplanar reconstruction (MPR) formatted images in a standard dose abdominal CT. Paper III, on the other hand, was an absolute assessment of image quality and pathology between the three phases of a CT Urography (CTU) protocol to explore the diagnostic value of low-dose abdominal CT. The anonymised images were displayed in random order and image quality was assessed by a group of radiologists using image quality criteria from the “European guidelines of quality criteria for CT”. The responses from the reviewer assessment were analysed statistically with ordinal logistic regression i.e. Visual Grading Regression (VGR).

Results in Paper I show that a small dose reduction (5-9 %) was possible using SAFIRE strength 1and indicated the need for further research to evaluate the dose reduction potential of higher strengths of the algorithm. In Paper II a 30% dose reduction was possible without change in ADMIRE algorithm strength as no improvement in image quality was observed between tube loads 98- and 140 mAs. When comparing tube loads 42 and 98 mAs, further dose reduction was possible with ADMIRE strength 3 (22-47%). However, for images reconstructed with ADMIRE strength 5, a dose reduction of 34-74% was possible for some, but not all image criteria. Image quality in low-contrast objects such as the liver parenchyma, was affected and a decline in diagnostic confidence was observed. Paper IV showed potential dose reductions are possible with increasing slice thickness from 1 mm to 2 mm (24-35%) and 1 mm to 3mm (25-41%). ADMIRE strength 3 continued to provide diagnostically acceptable images with possible dose reductions for all image criteria assessed. Despite objective evaluations showing a decrease in noise and an increase in contrast to noise ratio, ADMIRE strength 5 had diverse effects on the five image criteria, depending on slice thickness and further dose reductions were limited to certain image criteria. The findings do not support a general recommendation to replace ADMIRE3 with ADMIRE5 in clinical abdominal CT protocols.

Paper III studied another aspect of optimisation and results show that visualisation of renal anatomy was as expected in favour of the post-contrast phases when compared to the native phase. Assessment of pathology showed no significant differences between the three phases. Significantly higher diagnostic certainty for renal anatomy was observed for the post-contrast phases when compared to the native phase. Significantly high certainty scores were also seen for the nephrographic phase for incidental findings. The conclusion is that a low-dose series seems to be sufficient as a first-line modality in certain patient groups.

This thesis clinically evaluated the effect of IR in abdominal CT imaging and estimated potential dose reductions. The important conclusion from papers I, II and IV is that IR improves image quality in abdominal CT allowing for some dose reductions. However, the clinical utility of the highest strength of the algorithm is limited to certain criteria. The results can be used to optimise the clinical abdominal CT protocol. The conclusion from paper III may increase clinical awareness of the value of the low-dose abdominal protocol when choosing an imaging method for certain patient groups who are more sensitive to radiation.

Abstract [sv]

Datortomografi (DT) används i allt större omfattning vid bilddiagnostik och ger en viss stråldos till patienten. DT är en viktig, snabb och patientvänlig undersökningsteknik. En fördel med denna teknik är att bildmaterialet kan rekonstrueras i olika format för att åskådliggöra anatomin på bästasätt beroende på vilken frågeställning som ska besvaras. Joniserande strålning från dessa undersökningar anses öka risken för negativa effekter även om risken för den enskilde patient är mycket liten. Antalet datortomografiundersökningar ökar från år till år vilket kan leda till ökade stråldoser tillbefolkningen. Optimering av undersökningsteknik och val av undersökning för att minska negativa effekter av röntgenstrålning är därför nödvändig.

Det övergripande målet med avhandlingen var att utvärdera bildkvalitetvid en DT-undersökning av buken (då dessa medför en av de högstastråldoserna bland de vanliga röntgenundersökningarna), att kvantifieramöjlig stråldosminskning med hjälp av iterativa rekonstruktionsalgoritmer och att utvärdera diagnostiska värdet av lågdosundersökningsteknik vid DT-buk. Av de fyra delstudierna var delarbeten I, II och IV prospektiva och delarbete III retrospektivt.

För de prospektiva studierna, samlades bildmaterial in vid en kliniskberättigad undersökning av lågdos-DT av buken (delarbetet I), eller standarddos-DT av buken (delarbetet II och IV). Bilder rekonstruerades meden standard bildrekonstruktionsalgoritm, filtrerad återprojektion (FBP), och med styrka 1 av den iterativa algoritmen SAFIRE (delarbetet I). I delarbeten II och IV, gjordes bildrekonstruktioner med FBP och med styrka 3 och 5 av den iterativa algoritmen ADMIRE. Avidentifierade bildmaterialför varje patient visades parvis i slumpmässig ordning för ett antal granskare och bildkvaliteten bedömdes med hjälp av europeiska bildkriterier. I den retrospektiva studien, delarbete III, hämtades bildmaterialet från utförda DT-urografiundersökningar från bildarkivet. För varje undersökning visades bilder från varje fas i DT-urografiundersökningen separat i slumpmässig ordning. För samtliga delarbeten, hämtades bildkriteriernafrån ”European Guidelines of Quality Criteria for CT” och modifierades för att passa till varje studie. Granskarnas bedömning analyserades med ordinal logistisk regression så kallad visual grading regression (VGR). Resultat från delarbetet I visade att det fanns en signifikant inverkan av dos (p <0,001) och rekonstruktionsalgoritm (p <0,01) på samtliga bildkriterier, med en beräknad möjlig dosminskning på 5–9%. Delarbetet II visade att rekonstruktionsalgoritmen ADMIRE förbättrar bildkvaliteten i jämförelse med FBP. ADMIRE styrka 3 tillåter en dosminskning mellan 22–47% för samtliga bildkriterier medan ADMIRE styrka 5 tillåter en dosminskning mellan 34–74% för nästan alla bedömda bildkriterier utom återgivning av leverns parenkym. Ett mycket oväntat resultat var att bildkvalitén för 70% dosnivå bedömdes som högre eller likvärdig med 100% dosnivå, vilket innebar att stråldosen kan sänkas med 30% utan förändring i algoritm eller styrka.

Resultaten av delarbete III visade att avbildning av njuranatomi var som förväntat för varje fas med fördel för kontrastuppladdningsfaserna jämfört med den nativa fasen. Detta var inte ett oväntat resultat eftersom DT-urografiprotokollet är utformat för att visualisera njuranatomi på bästa möjliga sätt. Vid bedömning av patologiska fynd, erhölls dock små och ickesignifikanta skillnader mellan faserna. Däremot noterades signifikant högre bedömningssäkerhet för patologi i njurarna för de kontrast förstärkta faserna jämfört med nativfasen, och endast för bifynd signifikant högre poäng för parenkymfasen.

Delarbete IV visade att styrka 5 jämfört med styrka 3 av den iterativa rekonstruktionsalgoritmen, har olika effekter på bedömningen av bildkvalitetskriterierna. Ökning av MPR-snittjocklek från 1 mm till 2 mm eller 3mm, ger en förbättring i bildkvalité, vilket möjliggör en viss dosreduktion. Den kliniska användbarheten av ADMIRE styrka 5 är begränsad, medan ADMIRE styrka 3 levererar bättre bildkvalitet för samtliga undersökta bildkriterier vid datortomografiundersökning av buken.

Den viktigaste slutsatsen av delarbeten I, II och IV är att iterativa rekonstruktionsalgoritmer förbättrar bildkvalitet jämfört med FBP för samma stråldos och en dosminskning är möjlig. Detta kan användas för att optimera det kliniska DT-bukundersöknings protokoll. Slutsatsen för delarbetet III var att en lågdos-DT-bukundersökning är ett av många dosreduceringsalternativ, som möjligen kan användas för att minska strålningsbördan hos vissa patientgrupper som är mer känsliga för röntgenstrålning.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2019. p. 90
Series
Linköping University Medical Dissertations, ISSN 0345-0082 ; 1683
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
urn:nbn:se:liu:diva-160928 (URN)9789176850718 (ISBN)
Public defence
2019-11-07, Berzelius salen, Universitetssjukhuset, Linköping, 13:00 (English)
Opponent
Supervisors
Available from: 2019-10-15 Created: 2019-10-15 Last updated: 2019-10-15Bibliographically approved

Open Access in DiVA

fulltext(2444 kB)227 downloads
File information
File name FULLTEXT01.pdfFile size 2444 kBChecksum SHA-512
a49cb30cf6f954b6650e90f98fedcac361f1ba684f5c387b2938482e4f80a0d5e6cbef5acd77e49387f8435ddfb375ecf343073a4e5cc161e3f221e34b0a3717
Type fulltextMimetype application/pdf

Other links

Publisher's full textPubMedScopus

Authority records BETA

Kataria, BhartiSmedby, ÖrjanPersson, AndersSandborg, Michael

Search in DiVA

By author/editor
Kataria, BhartiNilsson Althen, JonasSmedby, ÖrjanPersson, AndersSökjer, HannibalSandborg, Michael
By organisation
Center for Medical Image Science and Visualization (CMIV)Division of Radiological SciencesDepartment of Radiology in LinköpingFaculty of Medicine and Health SciencesMedical radiation physicsDepartment of Medical and Health Sciences
In the same journal
European Radiology
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 227 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 302 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf