Optimization of the computational efficiency of a 3D, collapsed cone dose calculation algorithm for brachytherapy.
Linköpings universitet, Institutionen för medicin och hälsa, Medicinsk radiofysik. Linköpings universitet, Hälsouniversitetet. Östergötlands Läns Landsting, Kirurgi- och onkologicentrum, Radiofysikavdelningen.
Uppsala University.
2008 (Engelska)Ingår i: Medical physics (Lancaster), ISSN 0094-2405, Vol. 35, nr 4, s. 1611-1618Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

Brachytherapy dose calculations based on point kernel superposition using the collapsed cone method have been shown to accurately model the influence from finite dimensions of the patient and effects from heterogeneities including those of high atomic numbers. The collapsed cone method is for brachytherapy applications most effectively implemented through a successive-scattering approach, in which the dose from once and higher order of scattered photons is calculated separately and in successive scatter order. The calculation speed achievable is directly proportional to the number of directions used for point kernel discretization and to the number of voxels in the volume. In this work we investigate how to best divide the total number of directions between the two steps of successive-scattering dose calculations. Results show that the largest fraction of the total number of directions should be utilized in calculating the first-scatter dose. Also shown is how the number of directions required for keeping discretization artifacts at acceptably low levels decreases significantly in multiple-source configurations, as a result of the dose gradients being less steep than those around single sources. Investigating the number of kernel directions required to keep artifacts low enough within the high dose region of an implant (i.e., for dose levels above approximately 5%-10% of the mean central target dose) reveals similar figures for brachytherapy as for external beam applications, where collapsed cone superposition is clinically used. Also shown is that approximating point kernels with their isotropic average leads to small dose differences at low and intermediate energies, implying that the collapsed cone calculations can be done in a single operation common to all sources of the implant at these energies. The current findings show that collapsed cone calculations can be achieved for brachytherapy with the same efficiency as for external beams. This, combined with recent results on gains in efficiency through implementing the algorithm on graphical card parallel hardware indicates that dose can be calculated with account for heterogeneities and finite dimensions within a few seconds for large voxel arrays and is therefore of interest for practical application to treatment planning.

Ort, förlag, år, upplaga, sidor
2008. Vol. 35, nr 4, s. 1611-1618
Nyckelord [en]
brachytherapy, dose calculation, heterogeneities, collapsed cone, Monte Carlo
Nationell ämneskategori
Medicin och hälsovetenskap
Identifikatorer
URN: urn:nbn:se:liu:diva-19574DOI: 10.1118/1.2889777OAI: oai:DiVA.org:liu-19574DiVA, id: diva2:225493
Anmärkning
Original Publication: Asa Carlsson Tedgren and Anders Ahnesjö, Optimization of the computational efficiency of a 3D, collapsed cone dose calculation algorithm for brachytherapy., 2008, Medical physics (Lancaster), (35), 4, 1611-1618. http://dx.doi.org/10.1118/1.2889777 Copyright: American Institute of Physics http://www.aip.org/ Tillgänglig från: 2009-07-03 Skapad: 2009-06-29 Senast uppdaterad: 2017-12-13Bibliografiskt granskad

Open Access i DiVA

fulltext(1059 kB)588 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 1059 kBChecksumma SHA-512
885e6fef0faab4337b94048ff5ecd4161af804ab5485d62c23ebcb689800cdb4e2f63cbe856b3924d2a734629e0cb09ef993c4d1373be999c0f0781c95a27534
Typ fulltextMimetyp application/pdf

Övriga länkar

Förlagets fulltext

Personposter BETA

Carlsson Tedgren, Asa

Sök vidare i DiVA

Av författaren/redaktören
Carlsson Tedgren, Asa
Av organisationen
Medicinsk radiofysikHälsouniversitetetRadiofysikavdelningen
I samma tidskrift
Medical physics (Lancaster)
Medicin och hälsovetenskap

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 588 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

doi
urn-nbn

Altmetricpoäng

doi
urn-nbn
Totalt: 262 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf