4D Flow MRI-Based Pressure Loss Estimation in Stenotic Flows: Evaluation Using Numerical Simulations
Linköpings universitet, Institutionen för medicin och hälsa, Avdelningen för kardiovaskulär medicin. Linköpings universitet, Medicinska fakulteten.
Linköpings universitet, Institutionen för medicin och hälsa, Avdelningen för kardiovaskulär medicin. Linköpings universitet, Medicinska fakulteten.ORCID-id: 0000-0003-1942-7699
Linköpings universitet, Institutionen för medicin och hälsa, Avdelningen för kardiovaskulär medicin. Linköpings universitet, Medicinska fakulteten. Linköpings universitet, Centrum för medicinsk bildvetenskap och visualisering, CMIV. Region Östergötland, Hjärt- och Medicincentrum, Fysiologiska kliniken US.
Linköpings universitet, Institutionen för medicin och hälsa, Avdelningen för kardiovaskulär medicin. Linköpings universitet, Centrum för medicinsk bildvetenskap och visualisering, CMIV. Linköpings universitet, Institutionen för teknik och naturvetenskap, Medie- och Informationsteknik. Linköpings universitet, Medicinska fakulteten. Linköpings universitet, Tekniska fakulteten. Region Östergötland, Hjärt- och Medicincentrum, Fysiologiska kliniken US.ORCID-id: 0000-0003-1395-8296
2016 (Engelska)Ingår i: Magnetic Resonance in Medicine, ISSN 0740-3194, E-ISSN 1522-2594, Vol. 75, nr 4, s. 1808-1821Artikel i tidskrift (Refereegranskat) Published
Resurstyp
Text
Abstract [en]

Purpose: To assess how 4D flow MRI-based pressure and energy loss estimates correspond to net transstenotic pressure gradients (TPG(net)) and their dependence on spatial resolution. Methods: Numerical velocity data of stenotic flow were obtained from computational fluid dynamics (CFD) simulations in geometries with varying stenosis degrees, poststenotic diameters and flow rates. MRI measurements were simulated at different spatial resolutions. The simplified and extended Bernoulli equations, Pressure-Poisson equation (PPE), and integration of turbulent kinetic energy (TKE) and viscous dissipation were compared against the true TPG(net). Results: The simplified Bernoulli equation overestimated the true TPG(net) (8.74 +/- 0.67 versus 6.76 +/- 0.54 mmHg). The extended Bernoulli equation performed better (6.57 +/- 0.53 mmHg), although errors remained at low TPG(net). TPG(net) estimations using the PPE were always close to zero. Total TKE and viscous dissipation correlated strongly with TPG(net) for each geometry (r(2) > 0.93) and moderately considering all geometries (r(2) = 0.756 and r(2) = 0.776, respectively). TKE estimates were accurate and minorly impacted by resolution. Viscous dissipation was overall underestimated and resolution dependent. Conclusion: Several parameters overestimate or are not linearly related to TPG(net) and/or depend on spatial resolution. Considering idealized axisymmetric geometries and in absence of noise, TPG(net) was best estimated using the extended Bernoulli equation. (C) 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance.

Ort, förlag, år, upplaga, sidor
WILEY-BLACKWELL , 2016. Vol. 75, nr 4, s. 1808-1821
Nyckelord [en]
pressure loss; phase contrast magnetic resonance imaging; aortic valve disease; aortic coarctation
Nationell ämneskategori
Klinisk medicin
Identifikatorer
URN: urn:nbn:se:liu:diva-127426DOI: 10.1002/mrm.25772ISI: 000372910900043PubMedID: 26016805OAI: oai:DiVA.org:liu-127426DiVA, id: diva2:925262
Anmärkning

Funding Agencies|European Research Council [310612]; Swedish Research Council

Tillgänglig från: 2016-05-01 Skapad: 2016-04-26 Senast uppdaterad: 2018-10-10
Ingår i avhandling
1. Towards Personalized Models of the Cardiovascular System Using 4D Flow MRI
Öppna denna publikation i ny flik eller fönster >>Towards Personalized Models of the Cardiovascular System Using 4D Flow MRI
2018 (Engelska)Doktorsavhandling, sammanläggning (Övrigt vetenskapligt)
Abstract [en]

Current diagnostic tools for assessing cardiovascular disease mostly focus on measuring a given biomarker at a specific spatial location where an abnormality is suspected. However, as a result of the dynamic and complex nature of the cardiovascular system, the analysis of isolated biomarkers is generally not sufficient to characterize the pathological mechanisms behind a disease. Model-based approaches that integrate the mechanisms through which different components interact, and present possibilities for system-level analyses, give us a better picture of a patient’s overall health status.

One of the main goals of cardiovascular modelling is the development of personalized models based on clinical measurements. Recent years have seen remarkable advances in medical imaging and the use of personalized models is slowly becoming a reality. Modern imaging techniques can provide an unprecedented amount of anatomical and functional information about the heart and vessels. In this context, three-dimensional, three-directional, cine phase-contrast (PC) magnetic resonance imaging (MRI), commonly referred to as 4D Flow MRI, arises as a powerful tool for creating personalized models. 4D Flow MRI enables the measurement of time-resolved velocity information with volumetric coverage. Besides providing a rich dataset within a single acquisition, the technique permits retrospective analysis of the data at any location within the acquired volume.

This thesis focuses on improving subject-specific assessment of cardiovascular function through model-based analysis of 4D Flow MRI data. By using computational models, we aimed to provide mechanistic explanations of the underlying physiological processes, derive novel or improved hemodynamic markers, and estimate quantities that typically require invasive measurements. Paper I presents an evaluation of current markers of stenosis severity using advanced models to simulate flow through a stenosis. Paper II presents a framework to personalize a reduced-order, mechanistic model of the cardiovascular system using exclusively non-invasive measurements, including 4D Flow MRI data. The modelling approach can unravel a number of clinically relevant parameters from the input data, including those representing the contraction and relaxation patterns of the left ventricle, and provide estimations of the pressure-volume loop. In Paper III, this framework is applied to study cardiovascular function at rest and during stress conditions, and the capability of the model to infer load-independent measures of heart function based on the imaging data is demonstrated. Paper IV focuses on evaluating the reliability of the model parameters as a step towards translation of the model to the clinic.

Ort, förlag, år, upplaga, sidor
Linköping: Linköping University Electronic Press, 2018. s. 71
Serie
Linköping University Medical Dissertations, ISSN 0345-0082 ; 1641
Nyckelord
Magnetic resonance imaging, 4D Flow MRI, Computational modelling, Blood flow, Hemodynamics
Nationell ämneskategori
Biomedicinsk laboratorievetenskap/teknologi
Identifikatorer
urn:nbn:se:liu:diva-151751 (URN)10.3384/diss.diva-151751 (DOI)9789176852170 (ISBN)
Disputation
2018-11-05, Belladonna, Hus 511, Ingång 76, Våning 9, Campus US, Linköping, 13:00 (Engelska)
Opponent
Handledare
Forskningsfinansiär
EU, Europeiska forskningsrådet, 310612Vetenskapsrådet, 621-2014-6191Riksförbundet HjärtLung, 20140398
Tillgänglig från: 2018-10-05 Skapad: 2018-10-04 Senast uppdaterad: 2019-09-30Bibliografiskt granskad

Open Access i DiVA

fulltext(2041 kB)373 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 2041 kBChecksumma SHA-512
db1a997f9f1e2f4458b4d9b7f8ef8c007c4e22ba6b47252bece6b319b5bea66137748b0d36770d0b8ff11df878fe218a4f014394131deaa4c85cda5465368483
Typ fulltextMimetyp application/pdf

Övriga länkar

Förlagets fulltextPubMed

Person

Casas Garcia, BelénLantz, JonasDyverfeldt, PetterEbbers, Tino

Sök vidare i DiVA

Av författaren/redaktören
Casas Garcia, BelénLantz, JonasDyverfeldt, PetterEbbers, Tino
Av organisationen
Avdelningen för kardiovaskulär medicinMedicinska fakultetenCentrum för medicinsk bildvetenskap och visualisering, CMIVFysiologiska kliniken USMedie- och InformationsteknikTekniska fakulteten
I samma tidskrift
Magnetic Resonance in Medicine
Klinisk medicin

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 374 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

doi
pubmed
urn-nbn

Altmetricpoäng

doi
pubmed
urn-nbn
Totalt: 727 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf