The upgraded version of the GEM side-on thermal neutron detector was successfully tested in a neutron diffraction experiment on a reference sample using the INES diffractometer at the ISIS spallation neutron source, UK. The performance of the new (B4C)-B-10-based detector is compared to that of a standard He-3 tube, operating at the instrument as a part of the detectors assembly. The results show that the upgraded detector has a better resolution and an efficiency of the same order of magnitude of a He-3-based detector. Copyright (C) EPLA, 2018
Funding Agencies|e CNR (Italy)-STFC(UK) [06/20018]