Boron-10 lined RPCs for sub-millimeter resolution thermal neutron detectors: Feasibility study in a thermal neutron beamShow others and affiliations
2019 (English)In: Journal of Instrumentation, E-ISSN 1748-0221, Vol. 14, article id P01017Article in journal (Refereed) Published
Abstract [en]
The results of an experimental feasibility study of a position sensitive thermal neutron detector based on a resistive plate chamber (RPC) are presented. The detector prototype features a thin-gap (0.35 mm) hybrid RPC with an aluminium cathode and a float glass anode. The cathode is lined with a 2 mu m thick (B4C)-B-10 neutron converter enriched in B-10. A detection efficiency of 6.2% is measured at the neutron beam (lambda = 2.5 angstrom) for normal incidence. A spatial resolution better than 0.5 mm FWHM is demonstrated.
Place, publisher, year, edition, pages
IOP PUBLISHING LTD , 2019. Vol. 14, article id P01017
Keywords [en]
Neutron detectors (cold, thermal, fast neutrons); Resistive-plate chambers; Gaseous detectors
National Category
Accelerator Physics and Instrumentation
Identifiers
URN: urn:nbn:se:liu:diva-154132DOI: 10.1088/1748-0221/14/01/P01017ISI: 000455951500003OAI: oai:DiVA.org:liu-154132DiVA, id: diva2:1283574
Note
Funding Agencies|European Union [654000]; EU [676548]