Contrast Biases the Autocorrelation Phase Shift Estimation in Doppler Tissue Imaging
Linköping University, Department of Biomedical Engineering, Physiological Measurements. Linköping University, The Institute of Technology.
Lund University.
Linköping University, Department of Biomedical Engineering. Linköping University, The Institute of Technology.
Linköping University, Department of Biomedical Engineering, Physiological Measurements. Linköping University, The Institute of Technology.
Show others and affiliations
2009 (English)In: Ultrasound in Medicine and Biology, ISSN 0301-5629, Vol. 35, no 3, 447-457 p.Article in journal (Refereed) Published
Abstract [en]

Quantitative assessment of regional myocardial function at rest and during stress with Doppler tissue imaging (DTI) plays an important role in daily routine echocardiography. However, reliable visual analysis is largely dependent on image quality and adequate border delineation, which still remains a challenge in a significant number of patients. In this respect, an ultrasound contrast agent (UCA) is often used to improve visualization in patients with suboptimal image quality. The knowledge of how DTI measurements will be affected by UCA present in the tissue is therefore of significant importance for an accurate interpretation of local myocardial motion. The aim of this paper was to investigate how signal contribution from UCA and nonlinear wave propagation influence the performance of the autocorrelation phase shift estimator used for DTI applications. Our results are based on model experiments with a clinical 2-D grayscale scanner and computational simulations or the DTI velocity estimator for synthetically-derived pulses, simulated bubble echoes and experimentally-sampled RF data of transmitted pulses and backscattered contrast echoes. The results show that destruction of UCA present in the tissue will give rise to an apparent bidirectional velocity bias of individual velocity estimates, but that spatial averaging of individual velocity measurements within a region-of-interest will result in a negative bias (away from the transducer) of the estimated mean or mean peak velocity. The UCA destruction will also have a significant impact on the measured integrated mean velocity over time, i.e., displacement. To achieve improved visualization with UCA during DTI-examinations, we either recommend that it is performed at low acoustic powers, mechanical index <= 0.3, thereby minimizing the effects from bubble rupture, or that each Doppler pulse package is preceded by a destruction burst similar to "Flash imaging" to clear the target area of contrast microbubbles.

Place, publisher, year, edition, pages
2009. Vol. 35, no 3, 447-457 p.
Keyword [en]
Ultrasound, Tissue Doppler, Contrast, Microbubbles, Velocity estimation
National Category
Medical and Health Sciences
URN: urn:nbn:se:liu:diva-17277DOI: 10.1016/j.ultrasmedbio.2008.09.012OAI: diva2:208128
Available from: 2009-03-16 Created: 2009-03-16 Last updated: 2011-02-07
In thesis
1. On Nonlinear Acoustics in Contrast Echocardiography
Open this publication in new window or tab >>On Nonlinear Acoustics in Contrast Echocardiography
2010 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Ultrasound is one of the most commonly used noninvasive medical imaging techniques. Ultrasound contrast agents (UCA), consisting of encapsulated gas-filled microbubbles, have shown to increase the diagnostic precision in selected low echogenic patients. UCA also holds promise for bedside evaluation of myocardial perfusion quantification, but is not yet reproducible and specific enough for clinical use. In addition risks have been addressed when used, as first recommended, together with high mechanical index (MI) for reperfusion assessment by contrast destruction. We clinically observed increased myocardial velocities after UCA-administration when applied simultaneously with color tissue Doppler imaging (CTDI) arising the question if this increase was due to physiological factors or physical changes in the backscattered signals when UCA were present.

The aims of the thesis was to explain this velocity shift and simultaneously to contribute to a future safe and contrast specific application by further characterizing the non-linear acoustic properties of UCA when located in an acoustic field. Of specific interest was to evaluate in which way nonlinear wave propagation affects the response from UCA and if a change in pulse shape, length or polarity can be utilized to increase the nonlinear signal contribution.

Twelve patients with ischemic heart disease were examined with CTDI before and after UCA-administration in order to verify the change in peak systolic velocity. An experimental in vitro model including flow and tissue phantoms for UCA was established for CTDI. Raw data from single-element transducers and clinical ultrasound systems were collected for three different UCA and analyzed to determine if the observed velocity shift could be reproduced in vitro and to find a possible cause. Our results show in vivo and in vitro that UCA will affect the autocorrelation phase shift estimator used for CTDI in terms of contribution from rupturing UCA microbubbles, which explains the velocity shift. CTDI during contrast infusion should therefore be avoided unless it can be performed at low MI where the majority of the UCA are intact.

The computational model for spatial superposition of attenuated waves was modified to include an operator for pulse distortion from nonlinear wave propagation. The Matlab™ toolbox Bubblesim based on a modified Rayleigh-Plesset-equation and with insonation parameters such as frequency, pressure amplitude, pulse length and polarity was used to study the response from single microbubbles either for simulated pulses or for pulses generated by clinical ultrasound systems and single element transducers. The combination of the two models also provided a computational platform to asses pulse distortion from nonlinear wave propagation, the response of the UCA bubble and the linear backscatter of the low amplitude bubble echo. When evaluating the harmonic response in simulations and in vitro, the interaction of the excitation pulses with the contrast bubbles was identified as the main cause of nonlinear scattering, and a 2-3 dB increase of the second harmonic amplitude depends on nonlinear distortions of the incident pulse. By applying small changes of short (<3.5 cycles) and fragmented transmitted wideband pulses of 2-2.5 MHz, it is shown that inverted pulse polarity considerably modulates power without affecting a low and safe MI (<0.4), and the results lodged promise to further to enhance a contrast response.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2010. 114 p.
Linköping Studies in Science and Technology. Dissertations, ISSN 0345-7524 ; 1338
National Category
Engineering and Technology
urn:nbn:se:liu:diva-65418 (URN)978-91-7393-315-5 (ISBN)
Public defence
2010-11-12, Eken, Campus US, Linköpings universitet, Linököping, 09:15 (Swedish)
Available from: 2011-02-07 Created: 2011-02-07 Last updated: 2011-02-07Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full text

Search in DiVA

By author/editor
Ressner, MarcusAsk, PerJanerot Sjöberg, Birgitta
By organisation
Physiological MeasurementsThe Institute of TechnologyDepartment of Biomedical EngineeringClinical PhysiologyFaculty of Health SciencesDepartment of Clinical Physiology
In the same journal
Ultrasound in Medicine and Biology
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 498 hits
ReferencesLink to record
Permanent link

Direct link