Anatomic M-mode echocardiography: a new approach to assess regional myocardial function - A comparative in vivo and in vitro study of both fundamental and second harmonic imaging modes
Linköping University, Department of Medicine and Care, Clinical Physiology. Linköping University, Faculty of Health Sciences.
Show others and affiliations
1999 (English)In: Journal of the American Society of Echocardiography, ISSN 0894-7317, E-ISSN 1097-6795, Vol. 12, no 5, 300-307 p.Article in journal (Refereed) Published
Abstract [en]

Objective: To evaluate the accuracy of anatomic M-mode echocardiography (AMM).

Methods: Eight phantoms were rotated on a device at different insonation depths (IDs) in a water beaker. They were insonated with different transducer frequencies in fundamental imaging (FI) and second harmonic imaging (SHI), and the diameters were assessed with conventional M-mode echocardiography (CMM) and AMM with the applied angle correction (AC) after rotation. In addition, left ventricular wall dimensions were measured with CMM and AMM in FI and SHI in 10 volunteers.

Results: AC had the greatest effect on the measurement error in AMM followed by ID (AC: R2 = 0.295, ID: R2 = 0.268; P < .0001). SHI improved the accuracy, and a difference no longer existed between CMM and AMM with an AC up to 60 degrees. In vivo the limit of agreement between AMM and CMM was -1.7 to +1.8 mm in SHI.

Conclusion: Within its limitations (AC < 60 degrees; ID < 20 cm), AMM could be a robust tool in clinical practice.

Place, publisher, year, edition, pages
1999. Vol. 12, no 5, 300-307 p.
National Category
Medical and Health Sciences
URN: urn:nbn:se:liu:diva-26981DOI: 10.1016/S0894-7317(99)70050-7Local ID: 11616OAI: diva2:247532
Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2012-10-29
In thesis
1. Quantification of cardiovascular flow and motion: aspects of regional myocardial function and flow patterns in the aortic root and the aorta
Open this publication in new window or tab >>Quantification of cardiovascular flow and motion: aspects of regional myocardial function and flow patterns in the aortic root and the aorta
2004 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Quantification of cardiovascular flow and motion is essential in the diagnosis, treatment and follow-up of cardiovascular disease. The accuracy and quantification of many imaging methods used in this field have important shortfalls, however, that result from limitations in spatial and temporal dimensions. Improvement in application of these methods requires an in-depth understanding of the technical and perceptual aspects that contribute to errors in their use.

Visual assessment of echocardiographic images for asynchrony in regional myocardial motion during systolic contraction is an example of the need for better definition of limitations. The discernible delay in wall motion improved from 89 ms to 71 ms by allowing side-by-side comparison to normal motion. Clinically important delays are almost certainly missed with current "eyeballing" methods. Different and more quantitative approaches to this problem have been developed. Anatomic M-mode (AMM) assesses motion along an arbitrary line within a two-dimensional (2D) image, and was demonstrably robust in the clinical setting when used with second harmonic imaging at a depth less than 20 cm and with angle correction ofless than 60°. Doppler myocardial (DMI) imaging and strain rate imaging (SRI) were also shown to reliably demonstrate the effects of inotropic stimulation, total and severe ischemia on asynchrony in a closed chest pig model. Quantification of the changes induced by inotropy and total ischemia was possible with both methods, but the effects of stunning were not. Regional myocardial function and cardiovascular flow can also be assessed with time-resolved, three-directional, three-dimensional (3D) velocity data acquired using phase contrast magnetic resonance imaging (PC-MRI). This multidimensional data demonstrated longitudinal velocity gradients along all four walls of the left ventricle, with miuirnal apical longitudinal motion. The 3D velocity vector from single points in the ventricular wall shows that the motion over the cardiac cycle is complex in all dimensions. The flow patterns in the aortic root were also studied using time-resolved 3D PC-MRI in normal volunteers and patients who had undergone aortic-valve sparing surgery using straight Dacron grafts. In normals, vortices appeared in the sinuses of Valsalva in late systole, increased in size with the deceleration of aortic outflow and moved together as the valve closed in early diastole. These normal flow structures have never before been demonstrated in three dimensions in man. In the postoperative patients, lacking both sinuses and sinotubular junction, vortices were not observed.

Many imaging methods can be improved by a critical definition of the limits oftheir reliability. This can prompt the modifications and new methods which allow us to move beyond the original shortcomings and contribute new knowledge regarding the pathophysiology of cardiovascular disease.

Place, publisher, year, edition, pages
Linköping: Linköpings universitet, 2004. 63 p.
Linköping University Medical Dissertations, ISSN 0345-0082 ; 832
National Category
Medical and Health Sciences
urn:nbn:se:liu:diva-24209 (URN)3803 (Local ID)91-7373-804-2 (ISBN)3803 (Archive number)3803 (OAI)
Public defence
2004-02-20, Elsa Brändströmssalen, Universitetssjukhuset, Linköping, 13:00 (Swedish)
Available from: 2009-10-07 Created: 2009-10-07 Last updated: 2012-10-29Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full text

Search in DiVA

By author/editor
Escobar Kvitting, John-PederWranne, Bengt
By organisation
Clinical PhysiologyFaculty of Health Sciences
In the same journal
Journal of the American Society of Echocardiography
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 47 hits
ReferencesLink to record
Permanent link

Direct link