ESR dosimetry in the radiation therapy dose range: development of dosimetry systems and sensitive dosimeter materials
Linköping University, Department of Medicine and Care, Radio Physics. Linköping University, Department of Physics, Chemistry and Biology, Chemical Physics. Linköping University, Faculty of Health Sciences.
2001 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

ESR dosimetry with L-α-alanine as a dosimeter material is a well known and frequently used method for measuring high absorbed doses, for example at sterilisation and food irradiation. The increased sensitivity of modem spectrometers and an increased knowledge of the radiochemical properties of alanine have lowered the detection limit of ESR/alanine dosimetry to the dose range relevant for radiation therapy. The aim of this thesis is to contribute to an extended use of ESR dosimetry in radiation therapy, including development of an alanine-based gel dosimeter and investigation of alternative dosimeter materials.

The dosimeter material is often a polycrystalline powder, which can be mixed with a binder and formed to solid dosimeters in the shape of tablets, rods or films. It can also be distributed in a gel, to serve as both dosimeter material and phantom material. Thus, problems caused by the detector displacing the medium are avoided. The gel developed in this thesis is based on polycrystalline alanine, distributed in an agarose gel. It was tested for calibrated measurements in a brachytherapy situation. With further development, the alanine/agarose gel was found to have potential as a tool for verification of treatment plans, also at complicated dose distributions. One problem to be solved before the gel can be used at an arbitrary irradiation geometry is the low sensitivity of the gel. At present, the sensitivity puts a limit on the spatial resolution at low doses because of the high demands on measurement precision in radiation therapy.

The low sensitivity of the alanine/agarose gel, and also the need for a high precision at low doses together with a small dosimeter size when using traditional solid dosimeters, have raised the need for alternative, more sensitive, dosimeter materials than alanine. Such a material should form only one radiation induced radical, giving an ESR spectrum with few and narrow lines for easy evaluation. It should also be possible to increase the microwave power and the field modulation amplitude without severe distortion of the spectrum shape. Other important criteria are a low effective atomic number, a signal that is stable over time and, preferably, a linear dose response.

Two compounds are presented; ammonium tartrate and 2-methylalanine. Both  are more sensitive than alanine at low settings of microwave power and modulation amplitude, and fulfil the criteria of a low effective atomic number and a linear dose response. Ammonium tartrate is saturated already at low microwave power levels whereas the modulation amplitude can be increased to very high levels without saturation. However, the spectrum shape is then distorted. The signal-changes over time are rapid during the first hours after irradiation, but are then stabilised and can be corrected for.

The other investigated substance, 2-methylalanine, is more similar to alanine regarding signal stability over time and dependency of microwave power and field modulation amplitude. However, irradiation of 2-methylalanine results in only one detectable radical, where the alanine spectrum is composed of at least two radicals. The less complicated spectrum of 2-methylalanine makes it more safe to increase the microwave power and the modulation amplitude to their saturation levels.

Place, publisher, year, edition, pages
Linköping: Linköpings universitet , 2001. , 44 p.
Series
Linköping University Medical Dissertations, ISSN 0345-0082 ; 701
National Category
Medical and Health Sciences
Identifiers
URN: urn:nbn:se:liu:diva-27534Local ID: 12191ISBN: 91-7373-145-5 (print)OAI: oai:DiVA.org:liu-27534DiVA: diva2:248086
Public defence
2001-11-30, Berzeliussalen, Universitetssjukhuset, Linköping, 13:00 (Swedish)
Opponent
Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2012-10-17Bibliographically approved
List of papers
1. Dose response and fading characteristics of an Alanine-Agarose gel
Open this publication in new window or tab >>Dose response and fading characteristics of an Alanine-Agarose gel
1996 (English)In: Applied Radiation and Isotopes, ISSN 0969-8043, E-ISSN 1872-9800, Vol. 47, no 11-12, 1211-1217 p.Article in journal (Refereed) Published
Abstract [en]

The dose response of an alanine-agarose gel, analyzed by ESR spectrometry, and the stability of the radiation-induced free radicals have been investigated. The stability of the ESR signal is higher for dosimeter samples analyzed at 77 K than for dried samples, analyzed at room-temperature. The dose response is linear to within ±2% in the absorbed dose interval 2–100 Gy. The variations in spectral line shape were analyzed at temperatures between 77 and 270 K. The experimental ESR spectrum at 77 K was compared with a simulated spectrum of polycrystals of L-α-alanine.

National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-80203 (URN)10.1016/S0969-8043(96)00170-4 (DOI)
Available from: 2012-08-22 Created: 2012-08-22 Last updated: 2017-12-07Bibliographically approved
2. Radiation dose measurements with alanine/agarose gel and thin alanine films around a 192Ir brachytherapy source, using ESR spectroscopy
Open this publication in new window or tab >>Radiation dose measurements with alanine/agarose gel and thin alanine films around a 192Ir brachytherapy source, using ESR spectroscopy
Show others...
2002 (English)In: Physics in Medicine and Biology, ISSN 0031-9155, E-ISSN 1361-6560, Vol. 47, no 8, 1333-1356 p.Article in journal (Refereed) Published
Abstract [en]

Alanine/agarose gel and alanine films in stacks have been used for measurements of absorbed dose around an HDR 192Ir source in a vaginal cylinder-applicator, with and without a 180° tungsten shield. The gel and the films were analysed by means of ESR spectroscopy and calibrated against an ion chamber in a 4 MV photon beam to obtain absolute dose values. The gel serves as both dosimeter and phantom material, and the thin (130 μm) films are used to achieve an improved spatial resolution in the dose estimations.

Experimental values were compared with Monte Carlo simulations using two different codes. Results from the measurements generally agree with the simulations to within 5%, for both the alanine/agarose gel and the alanine films.

National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-27010 (URN)10.1088/0031-9155/47/8/308 (DOI)11651 (Local ID)11651 (Archive number)11651 (OAI)
Note
On the day of the defence day the status of this article was submitted.Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2017-12-13Bibliographically approved
3. Ammonium tartrate as an ESR dosimeter material
Open this publication in new window or tab >>Ammonium tartrate as an ESR dosimeter material
Show others...
1999 (English)In: Applied Radiation and Isotopes, ISSN 0969-8043, E-ISSN 1872-9800, Vol. 50, no 5, 955-965 p.Article in journal (Refereed) Published
Abstract [en]

This study is one step in the search for an ESR dosimeter material with a higher signal intensity than the commonly used l-α-alanine, to be useful in the clinical dose range (approximately 0.1–20 Gy). The substance ammonium tartrate was found and investigated regarding signal intensity, radical stability, dose response and dose resolution. The ESR signal intensity of ammonium tartrate was shown to be more than twice the intensity of the alanine signal. The data indicate that an unstable radiation induced radical contributes to the ESR signal initially; after a couple of hours it has converted to a secondary radical which has a decay slow enough to be considered stable during the first two weeks after irradiation. Ammonium tartrate has a linear dose response in the investigated range of 0.5–4000 Gy and a dose resolution of 0.1 Gy at the 0.5 Gy level where, as a comparison, the corresponding value for alanine is 0.3 Gy. We thus find the substance suitable for clinical dosimetry.

National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-26760 (URN)10.1016/S0969-8043(98)00120-1 (DOI)11360 (Local ID)11360 (Archive number)11360 (OAI)
Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2017-12-13Bibliographically approved
4. Development of ammonium tartrate as an ESR dosimeter material for clinical purposes
Open this publication in new window or tab >>Development of ammonium tartrate as an ESR dosimeter material for clinical purposes
2000 (English)In: Applied Radiation and Isotopes, ISSN 0969-8043, E-ISSN 1872-9800, Vol. 52, no 5, 1235-1241 p.Article in journal (Refereed) Published
Abstract [en]

The crystalline substance ammonium tartrate is investigated with respect to its suitability as a clinical dosimetry material. The properties investigated are: the radical stability for absorbed doses relevant for clinical use, the improvement in sensitivity when the crystals are deuterated and the linear electron tranfer (LET)-dependence. After photon irradiation to an absorbed dose of 20 Gy, the signal increases rapidly during the first 6 h. After this period, the changes are more moderate and can be corrected for. The signal-to-noise ratio of irradiated ammonium tartrate is twice the corresponding value of alanine. By deuterating the crystals, the sensitivity can be further improved by a factor of 1.4. As expected, the signal decreases with increasing LET of the radiation, but no changes in the spectrum shape was observed.

National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-27009 (URN)10.1016/S0969-8043(00)00077-4 (DOI)11650 (Local ID)11650 (Archive number)11650 (OAI)
Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2017-12-13Bibliographically approved
5. EPR Dosimetric Properties of 2-Methylalanine: EPR, ENDOR and FT-EPR Investigations
Open this publication in new window or tab >>EPR Dosimetric Properties of 2-Methylalanine: EPR, ENDOR and FT-EPR Investigations
2002 (English)In: Radiation Research, ISSN 0033-7587, E-ISSN 1938-5404, Vol. 157, no 2, 113-121 p.Article in journal (Refereed) Published
Abstract [en]

To find an EPR dosimeter material that is sensitive enough for clinical use, the substance 2-methylalanine (2MA) with the chemical structure (CH3)2C(NH3+)COO was tested for its sensitivity to ionizing radiation, dose response, and radical stability over time. At equal and moderate settings of microwave power and modulation amplitude, 2MA was found to be 70% more sensitive than l-α-alanine, which is the most common EPR dosimeter material today. The dose response is linear, at least in the dose range of interest (0.5–100 Gy), and the time-dependent variations in signal intensity are very small and may be corrected for easily. The energy dependence of the stopping power and energy absorption was calculated and was found to be similar to that of alanine. The dependence of the signal intensity on microwave power and modulation amplitude was investigated, and the optimal settings were found to be 25 mW (Bruker ER 4102ST) and 12 gauss, respectively. Single crystals of 2MA were analyzed using ENDOR and ENDOR-induced EPR to identify the radiation-induced radicals that formed. Only one radical, in which the amino group is detached from the original molecule, was identified. This radical is obviously dominating and is apparently the only one relevant for dosimetry purposes. The complete set of coupling parameters for three hyperfine couplings is reported. The power saturation properties and spectral line width are ruled by the relaxation times T1 and T2. To determine the relaxation times of 2MA, pulsed EPR experiments were performed on single crystals. Two different values of T1 were obtained, one in the range 1–3 µs, shown to be of importance for the dosimetry properties, and another that is strongly anisotropic with a value between 10 and 35 µs that does not seem to affect the saturation behavior. T2 was estimated to be of the order of 200–300 ns.

National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-27098 (URN)10.1667/0033-7587(2002)157[0113:EDPOME]2.0.CO;2 (DOI)11745 (Local ID)11745 (Archive number)11745 (OAI)
Available from: 2009-10-08 Created: 2009-10-08 Last updated: 2017-12-13Bibliographically approved

Open Access in DiVA

No full text

Authority records BETA

Olsson, Sara

Search in DiVA

By author/editor
Olsson, Sara
By organisation
Radio PhysicsChemical PhysicsFaculty of Health Sciences
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar

isbn
urn-nbn

Altmetric score

isbn
urn-nbn
Total: 133 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf