liu.seSearch for publications in DiVA
Change search
ReferencesLink to record
Permanent link

Direct link
The impression technique for assessment of tissue oedema: instrumentation, evaluation and applications
Linköping University, Department of Biomedical Engineering. Linköping University, The Institute of Technology.
1993 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

An instrument for clinical noninvasive assessment of tissue oedema based on an impression method was developed. The method measures and evaluates the decaying force, due to translocation of tissue fluid, during mechanical compression of any site of tissue. We applied the impression method on physical models, animal models, and patients. Significant parameters for the assessment of tissue oedema that estimated tissue fluid translocation and tissue pressure could be derived from the registered impression force curves.

Accuracy was determined theoretically and reproducibility was estimated on plastic foam. We described the clinical procedure for the instrument, and preliminary results from patients with chronic pitting oedema showed that the instrument detected larger fluid translocation on oedematous sites than on non-oedematous sites. We concluded that the instrument was acceptable for accurate measurements on biological tissue.

Evaluation was performed in a rat testis model in which testicular interstitial fluid volume could be changed both artificially by 30-min infusions of different fluids with different fluid resistance properties, and pharmacologically by administration of hormones. We found that tissue pressure increased with infused fluid volume, and changes as small as 16 μl (7 % of total testis interstitial fluid volume) could be detected. Fluid translocation changed depending upon the infused fluid's resistance properties. Hormone-induced changes in rat-testis oedema altered both fluid translocation and tissue pressure. Discrete changes in vascular permeability were monitored.

Investigation of generalised oedema in patients suffering from burn injury showed that tissue fluid translocation increased up to a maximum value after 6 days postburn and declined thereafter. We found tissue pressure to be relatively high during the first 7 days postburn as compared with 3-week postburn values. Force curve analysis suggested a flux of water-like fluid from the vasculature to the interstitial space during the first 6 days postburn. The course of postburn tissue swelling could be followed and estimated with the impression technique.

Comparison with a new tactile sensor that measured physical properties of soft tissue showed that both methods detected changes in silicone hardness/softness and in hormone-induced changes of rat-testis interstitial fluid. We concluded that impression force estimated hardness of soft tissue, which can be helpful when investigating hardness of oedematous tissue.

Place, publisher, year, edition, pages
Linköping: LJ Foto & Montage , 1993. , 51 p.
Linköping Studies in Science and Technology. Dissertations, ISSN 0345-7524 ; 309
, Umeå University Medical Dissertations New Series, ISSN 0346-6612 ; 372
National Category
Medical and Health Sciences
URN: urn:nbn:se:liu:diva-29458Local ID: 14806ISBN: 91-7871-097-9OAI: diva2:250273
Public defence
1993-05-28, Röntgens föreläsningssal, Norrlands universitetssjukhus, Umeå, 09:15 (Swedish)

Papers, included in the Ph.D. thesis, are not registered and included in the posts from 1999 and backwards.

Available from: 2009-10-09 Created: 2009-10-09 Last updated: 2013-01-16

Open Access in DiVA

No full text

By organisation
Department of Biomedical EngineeringThe Institute of Technology
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Total: 294 hits
ReferencesLink to record
Permanent link

Direct link