liu.seSearch for publications in DiVA
Change search
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf
Calculation and analysis of DQE for some image detectors in mammography
Linköping University, Department of Medicine and Care, Radio Physics. Linköping University, Faculty of Health Sciences. Östergötlands Läns Landsting, Centre of Surgery and Oncology, Department of Radiation Physics.ORCID iD: 0000-0003-3352-8330
1998 (English)Report (Other academic)
Abstract [en]

The development and clinical testing of digital detector designs for mammography are making rapid progress and there is widespread interest in comparing the performance of these new detectors to that of analogue screen-film mammography systems. In this report, Monte Carlo calculations of the x-ray absorption characteristics (single-event distribution), the quantum absorption, Aq, and detective quantum, DQEq, efficiencies are made and compared to results from the literature. Detectors of CsI and Si of various thicknesses are compared to a state-ofthe art analogue, screen-film system (Gd2O2S) in the energy range 1-35 keV. The results show that 1.5 mm thick Si detectors will have the same DQEq as commonly used Gd2O2S fluorescent screens and that a CsI phosphor of 80 mm has similar DQEq as 1.0 mm Si. The total DQE (including added noise and inherent detector unsharpness) of fluorescent screen-film systems will be significantly reduced from this value due to the light scatter, film noise and the inherent limitations caused by the film characteristic curve. This indicates that also thinner Si detectors (0.3-0.5 mm), which do not suffer from these limitations but from a comparably low Aq may have a total performance DQE(f) comparable to that of traditional screen-film based, image detectors.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press , 1998. , 20 p.
Series
Report / Department of Radiology, Linköping University, 1998-, 86
Keyword [en]
digital mammography, detective quantum efficiency (DQE), signal-tonoise ratio (SNR), Monte Carlo calculations, silicon detectors
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:liu:diva-57821ISRN: LIU-RAD-R-086OAI: oai:DiVA.org:liu-57821DiVA: diva2:327851
Available from: 2010-06-30 Created: 2010-06-30 Last updated: 2013-09-03Bibliographically approved

Open Access in DiVA

Calculation and analysis of DQE for some image detectors in mammography(314 kB)1738 downloads
File information
File name FULLTEXT02.pdfFile size 314 kBChecksum SHA-512
374ee2cc67fc65c8fc2c4b18ec2db06b140faf65210778edc184efc86a03670df7090a1763cd4e1efaf8e32e750efb0e2df346c68a34d3c1d208a78afe730eb3
Type fulltextMimetype application/pdf

Authority records BETA

Sandborg, Michael

Search in DiVA

By author/editor
Sandborg, Michael
By organisation
Radio PhysicsFaculty of Health SciencesDepartment of Radiation Physics
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 1738 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 106 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf