Calibration of alanine dosimeters
Linköping University, Department of Medicine and Care, Radio Physics. Linköping University, Faculty of Health Sciences. Östergötlands Läns Landsting, Centre of Surgery and Oncology, Department of Radiation Physics.
Östergötlands Läns Landsting, Centre of Surgery and Oncology, Department of Radiation Physics.
2001 (English)Report (Other academic)
Abstract [en]

In all kinds of radiation therapy it is of crucial importance to thoroughly determine the absorbed dose in the tumour, in surrounding normal tissue and in risk organs. This is done using various types of dosimeters, all with their advantages and disadvantages for different situations.

Measurements of the dose distribution in a volume in terms of absolute absorbed dose are needed, for example to verify Monte Carlo calculations or clinical dose plans experimentally. To make such absolute dose measurements possible, the dosimeter requires calibration towards a dosimeter with a calibration factor that is traceable to a primary standard dosimetry laboratory. To not put the outcome of a radiation treatment at risk, the uncertainty in the absorbed dose determination at points of interest must not be higher than 5% (1 standard deviation) according to the ICRU (1976).

The aim of this report is to indicate a way of calibrating two types of L-a-alanine dosimeters, an alanine/agarose gel and thin alanine films. The influence of some factors on the ESR signal from the alanine dosimeters is investigated, and suggestions are made on how to take these factors into account.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press , 2001. , 25 p.
Report / Department of Radiology, Linköping University, 1998-, 92a
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:liu:diva-57833ISRN: LiU-RAD-R--92a--SEOAI: diva2:327962
Available from: 2010-07-01 Created: 2010-07-01 Last updated: 2010-07-01Bibliographically approved

Open Access in DiVA

Calibration of alanine dosimeters(138 kB)454 downloads
File information
File name FULLTEXT01.pdfFile size 138 kBChecksum SHA-512
Type fulltextMimetype application/pdf

Search in DiVA

By author/editor
Olsson, Sara
By organisation
Radio PhysicsFaculty of Health SciencesDepartment of Radiation Physics
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 454 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Total: 57 hits
ReferencesLink to record
Permanent link

Direct link