Increased accuracy of echocardiographic measurement of flow using automated spherical integration of multiple plane velocity vectors
Linköping University, Department of Biomedical Engineering. Linköping University, The Institute of Technology.
Linköping University, Department of Medicine and Care, Clinical Physiology. Linköping University, Faculty of Health Sciences.
Linköping University, Department of Biomedical Engineering. Linköping University, The Institute of Technology.
1999 (English)In: Ultrasound in Medicine and Biology, ISSN 0301-5629, Vol. 25, no 2, 249-257 p.Article in journal (Refereed) Published
Abstract [en]

The calculation of blood flow in the heart by surface integration of velocity vectors (SIVV) using Doppler ultrasound is independent of the angle. Flow is normally calculated from velocity in a spherical thick shell with its center located at the ultrasound transducer. In a numerical simulation, we have shown that the ratio between minor and major axes of an elliptic flow area substantially influences the accuracy of the estimation of flow in a single scan plane. The accuracy of flow measurements by SIVV can be improved by calculating the mean of the values from more than one scan plane. We have produced an automated computer program that includes an antialiasing procedure. We confirmed an improvement of flow measurements in a pulsatile hydraulic flow model, the 95% confidence interval for single estimations being reduced from 20% to 10% (p < 0.05) using the newly developed software. We think that the SIVV method has important implications for clinical transthoracic echocardiography.

Place, publisher, year, edition, pages
1999. Vol. 25, no 2, 249-257 p.
National Category
Medical Laboratory and Measurements Technologies
URN: urn:nbn:se:liu:diva-67263DOI: 10.1016/S0301-5629(98)00159-8PubMedID: 10320314OAI: diva2:408924
Available from: 2011-04-06 Created: 2011-04-06 Last updated: 2015-04-09
In thesis
1. On cardiac flow quantification with ultrasound colour doppler
Open this publication in new window or tab >>On cardiac flow quantification with ultrasound colour doppler
2000 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

This thesis deals with the estimation of blood flow in the heart and larger vessels where control-volume methods are applied using ultrasound Doppler technique. In particular two control-volume techniques were investigated: The proximal isovelocity surface area method, (PISA) and the Surface Integration of Velocity Vectors method, (SIVV).

For PISA, computational fluid dynamics, (CFD) was used for non-stationary flow and non-planar circular geometries where special emphasis was given to the influence from the angle of the valvular leaflets on the proximal surface area. The CFD results were compared with ultrasound measurements, in an in-vitro model with controlled geometry and flow characteristics. Three different valvular geometries were used: planar, reversed cone and funnel. In these idealised CFD and experimental models it was found that there is support to use the hemispherical PISA approach for the geometries investigated provided that the flow is not to high in the reversed cone and funnel case. At high flows the actual proximal geometry should be used instead of an entire hemisphere.

A hydraulic pulsatile model was used in developing a platform with in-house software where the SIVV flows automatically may be calculated from a digitally stored raw data. An antialiasing algorithm was developed to allow for measurement of aliased data in order to increase the dynamic velocity range. The antialiasing algorithm was found to improve the estimation of SIVV flow.

The influence on the flow estimate was investigated with respect to the number of scan-planes using a numerical model and in-vitro and in-vivo model experiments. It was found that a minimum of two scan-planes are needed when flow conditions and geometry is close to circular, otherwise the recommendation is four scan-planes.

A steady state and a pulsatile model was used to evaluate accuracy of the SIVV method more extensively in vitro. SIVV was found to be accurate and repeatable with a slight underestimation in the pulsatile model but within the ±10% range. In the steady state model a strong correlation was found between SIVV and timed flow. However, since discrepancies in regression equations were obtained for different tube diameters further investigation of steady state flows in vessels of small diameter are needed.

An in-vivo model was designed to study the possibility to use the SIVV method to measure cardiac output in a paediatric model in haemodynamically unstable subjects and to investigate what measurement site to use. Epicardial measurements were performed on a series of piglets using two different temporal resolutions. SJVV accuracy was compared with ultrasound transit time flow and was found to be in parity or better than current invasive methods. Inter- and lntraobserver variability was found to be low.

Place, publisher, year, edition, pages
Linköping: Linköpings universitet, 2000. 48 p.
Linköping Studies in Science and Technology. Dissertations, ISSN 0345-7524 ; 625
National Category
Medical and Health Sciences
urn:nbn:se:liu:diva-29447 (URN)14794 (Local ID)91-7219-702-1 (ISBN)14794 (Archive number)14794 (OAI)
Public defence
2000-04-17, Berzeliussalen, Universitetssjukhuset, Linköping, 09:15 (Swedish)
Available from: 2009-10-09 Created: 2009-10-09 Last updated: 2013-02-25

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Brandberg, JoakimJanerot Sjöberg, BirgittaAsk, Per
By organisation
Department of Biomedical EngineeringThe Institute of TechnologyClinical PhysiologyFaculty of Health Sciences
In the same journal
Ultrasound in Medicine and Biology
Medical Laboratory and Measurements Technologies

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 105 hits
ReferencesLink to record
Permanent link

Direct link