liu.seSearch for publications in DiVA
Endre søk
Link to record
Permanent link

Direct link
Svensson, Anders
Publikasjoner (2 av 2) Visa alla publikasjoner
Svensson, A., Kovesdy, C. P., Kvitting, J.-P. E., Holm, J. & Zoltán, S. (2025). Changes in urinary pO2, pCO2 and lactate are associated with acute kidney injury during cardiac surgery. In: : . Paper presented at Scandinavian Association of Thoracic Surgery, SATS meeting, Reykjavik, Island 2025.
Åpne denne publikasjonen i ny fane eller vindu >>Changes in urinary pO2, pCO2 and lactate are associated with acute kidney injury during cardiac surgery
Vise andre…
2025 (engelsk)Konferansepaper, Poster (with or without abstract) (Annet vitenskapelig)
HSV kategori
Identifikatorer
urn:nbn:se:liu:diva-222070 (URN)
Konferanse
Scandinavian Association of Thoracic Surgery, SATS meeting, Reykjavik, Island 2025
Tilgjengelig fra: 2026-03-19 Laget: 2026-03-19 Sist oppdatert: 2026-03-26
Ohlsson, L., Sandstedt, M., Papageorgiou, J.-M., Svensson, A., Bolger, A. F., Tamás, É., . . . Lantz, J. (2024). Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.. European heart journal. Imaging methods and practice, 2(3), Article ID qyae082.
Åpne denne publikasjonen i ny fane eller vindu >>Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.
Vise andre…
2024 (engelsk)Inngår i: European heart journal. Imaging methods and practice, ISSN 2755-9637, Vol. 2, nr 3, artikkel-id qyae082Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]

AIMS: The HeartMate 3 (HM3) implantable left ventricular assist device connects the left ventricle apex to the aorta via an outflow graft. Extrinsic obstruction of the graft (eOGO) is associated with serious morbidity and mortality and recently led to a Food and Drug Administration Class 1 device recall of HM3. This study aimed to provide a better understanding of the haemodynamic impact of extrinsic stenoses.

METHODS AND RESULTS: Computed tomography (CT) images of two retrospectively identified patients with eOGO (29 and 36% decrease in cross-sectional area, respectively, by radiological evaluation) were acquired with a novel photon-counting CT system. Numerical evaluations of haemodynamics were conducted using a high-fidelity 3D computational fluid dynamics approach on both the patient-specific graft geometries and in two virtually augmented stenotic severities and three device flows. Visual analysis identified increased velocity, pressure, and turbulent flow in the outer anterior curvature of the outflow graft; however, changes in graft pressure gradients were slight (1-9 mmHg) across the range of stenosis severities and flow rates tested.

CONCLUSION: Evidence of eOGO during HM3 support and the recent device recall can provoke clinical apprehension and interventions. The haemodynamic impact of a stenosis detected visually or by quantification of cross-sectional area reduction may be difficult to predict and easily overestimated. This numerical study suggests that, for clinically encountered flow rates and stenosis severities below 61% in cross-sectional area decrease, eOGO may have low haemodynamic impact. This suggests that patients without symptoms or signs consistent with haemodynamically significant obstruction might be managed expectantly.

sted, utgiver, år, opplag, sider
Oxford University Press, 2024
Emneord
CFD, HeartMate 3, LVAD, haemodynamics, photon-counting CT
HSV kategori
Identifikatorer
urn:nbn:se:liu:diva-208765 (URN)10.1093/ehjimp/qyae082 (DOI)39224624 (PubMedID)
Tilgjengelig fra: 2024-10-23 Laget: 2024-10-23 Sist oppdatert: 2025-10-02bibliografisk kontrollert
Organisasjoner