liu.seSök publikationer i DiVA
Ändra sökning
Länk till posten
Permanent länk

Direktlänk
Svensson, Anders
Publikationer (2 of 2) Visa alla publikationer
Svensson, A., Kovesdy, C. P., Kvitting, J.-P. E., Holm, J. & Zoltán, S. (2025). Changes in urinary pO2, pCO2 and lactate are associated with acute kidney injury during cardiac surgery. In: : . Paper presented at Scandinavian Association of Thoracic Surgery, SATS meeting, Reykjavik, Island 2025.
Öppna denna publikation i ny flik eller fönster >>Changes in urinary pO2, pCO2 and lactate are associated with acute kidney injury during cardiac surgery
Visa övriga...
2025 (Engelska)Konferensbidrag, Poster (med eller utan abstract) (Övrigt vetenskapligt)
Nationell ämneskategori
Kirurgi
Identifikatorer
urn:nbn:se:liu:diva-222070 (URN)
Konferens
Scandinavian Association of Thoracic Surgery, SATS meeting, Reykjavik, Island 2025
Tillgänglig från: 2026-03-19 Skapad: 2026-03-19 Senast uppdaterad: 2026-03-26
Ohlsson, L., Sandstedt, M., Papageorgiou, J.-M., Svensson, A., Bolger, A. F., Tamás, É., . . . Lantz, J. (2024). Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.. European heart journal. Imaging methods and practice, 2(3), Article ID qyae082.
Öppna denna publikation i ny flik eller fönster >>Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.
Visa övriga...
2024 (Engelska)Ingår i: European heart journal. Imaging methods and practice, ISSN 2755-9637, Vol. 2, nr 3, artikel-id qyae082Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

AIMS: The HeartMate 3 (HM3) implantable left ventricular assist device connects the left ventricle apex to the aorta via an outflow graft. Extrinsic obstruction of the graft (eOGO) is associated with serious morbidity and mortality and recently led to a Food and Drug Administration Class 1 device recall of HM3. This study aimed to provide a better understanding of the haemodynamic impact of extrinsic stenoses.

METHODS AND RESULTS: Computed tomography (CT) images of two retrospectively identified patients with eOGO (29 and 36% decrease in cross-sectional area, respectively, by radiological evaluation) were acquired with a novel photon-counting CT system. Numerical evaluations of haemodynamics were conducted using a high-fidelity 3D computational fluid dynamics approach on both the patient-specific graft geometries and in two virtually augmented stenotic severities and three device flows. Visual analysis identified increased velocity, pressure, and turbulent flow in the outer anterior curvature of the outflow graft; however, changes in graft pressure gradients were slight (1-9 mmHg) across the range of stenosis severities and flow rates tested.

CONCLUSION: Evidence of eOGO during HM3 support and the recent device recall can provoke clinical apprehension and interventions. The haemodynamic impact of a stenosis detected visually or by quantification of cross-sectional area reduction may be difficult to predict and easily overestimated. This numerical study suggests that, for clinically encountered flow rates and stenosis severities below 61% in cross-sectional area decrease, eOGO may have low haemodynamic impact. This suggests that patients without symptoms or signs consistent with haemodynamically significant obstruction might be managed expectantly.

Ort, förlag, år, upplaga, sidor
Oxford University Press, 2024
Nyckelord
CFD, HeartMate 3, LVAD, haemodynamics, photon-counting CT