Increasing the dose efficiency in silicon photon-counting detectors utilizing dual shapersVisa övriga samt affilieringar
2018 (Engelska)Ingår i: PROCEEDINGS VOLUME 10573 SPIE MEDICAL IMAGING, 10-15 FEBRUARY 2018 Medical Imaging 2018: Physics of Medical Imaging, SPIE - International Society for Optical Engineering, 2018, Vol. 10573, artikel-id UNSP 105734WKonferensbidrag, Publicerat paper (Refereegranskat)
Abstract [en]
Silicon photon-counting spectral detectors are candidates for the next generation of medical CT. For silicon detectors, a low noise floor is necessary to obtain good detection efficiency. A low noise floor can be obtained by having a slow shaping filter in the ASIC, but this leads to a long dead-time, thus decreasing the count-rate performance. In this work, we evaluate the benefit of utilizing two sub-channels with different shaping times. It is shown by simulation that utilizing a dual shaper can increase the dose efficiency for equal count-rate capability by up to 17%.
Ort, förlag, år, upplaga, sidor
SPIE - International Society for Optical Engineering, 2018. Vol. 10573, artikel-id UNSP 105734W
Serie
Progress in Biomedical Optics and Imaging, ISSN 1605-7422 ; 10573
Nyckelord [en]
dose efficiency; silicon detector; shaping time; ASIC; photon-counting
Nationell ämneskategori
Atom- och molekylfysik och optik
Identifikatorer
URN: urn:nbn:se:liu:diva-151532DOI: 10.1117/12.2293453ISI: 000436173700167ISBN: 9781510616363 (tryckt)OAI: oai:DiVA.org:liu-151532DiVA, id: diva2:1250585
Konferens
SPIE Medical Imaging, 10-15 February, 2018, Houston, Texas, United States