Pinloc or Hansson pins: a multicenter, randomized controlled study of 439 patients treated for femoral neck fractures
2023 (English)In: OTA international : the open access journal of orthopaedic trauma, ISSN 2574-2167, Vol. 6, no 4, article id e282Article in journal (Refereed) Published
Abstract [en]
Objectives: To compare the recently developed Hansson Pinloc system, which features 3 cylindrical parallel pins with hooks connected through a fixed-angle interlocking plate, with the Hansson Pin System (2 hook pins) for the treatment of femoral neck fractures.
Design: One hundred fourteen patients with displaced femoral neck fractures and 325 patients with nondisplaced fractures from 9 orthopaedic centers were randomized to either Hansson Pinloc system or Hansson Pin System and followed for 2 years or until death. Age at inclusion was 50 years or older.
Main outcome measurements: The primary outcome was failure (defined as early displacement, nonunion, symptomatic avascular necrosis, or deep infection). Secondary outcomes included revision surgery, Timed Up and Go (TUG) test and patient-reported outcome measures (PROMs: EQ-5D and WOMAC).
Results: For nondisplaced fractures, the incidence of failure was 14% (23/169) in the Pinloc group and 16% (25/156) in the Hansson group. For displaced fractures, the analysis was stratified by age. Patients aged 50-69 years with displaced fractures showed a 2-year failure rate of 44% (17/39) in the Pinloc group versus 44% (16/36) in the Hansson group. For patients 70 years or older with displaced fractures, 33% (7/21) in the Pinloc group versus 22% (4/18) in the Hansson group failed. At 3 and 12 months, no clinically significant differences between treatment groups were found for EQ-5D-3L, WOMAC, or for the TUG in any fracture type or age group.
Conclusions: There were no advantages for Pinloc in any of the studies aspects.
Level of evidence: 1.
Place, publisher, year, edition, pages
Wolters Kluwer Health , 2023. Vol. 6, no 4, article id e282
Keywords [en]
Hansson Pin System; Hansson Pinloc system; femoral neck fracture; internal fixation
National Category
Embedded Systems
Identifiers
URN: urn:nbn:se:liu:diva-202790DOI: 10.1097/OI9.0000000000000282PubMedID: 37744995OAI: oai:DiVA.org:liu-202790DiVA, id: diva2:1853510
2024-04-222024-04-222024-04-22