Purpose: Accurate measurements of trabecular bone microarchitecture are required for a proper assessment of bone fragility. Photon-counting detector CT (PCD-CT) has different technical properties than conventional CT, resulting in higher resolution and thereby potentially enabling in-vivo measurement of trabecular microarchitecture. The purpose of this study was to quantify trabecular bone microarchitectural parameters with PCDCT at varying radiation doses and compare this to mu CT as gold standard. Method: Both distal radii, distal tibiae, femoral heads, and two vertebrae were dissected from one human. All specimens were scanned ex-vivo on a PCD-CT system (slice increment 0.1 mm; pixel size 0.1042-0.127 mm) and a mu CT system (isotropic voxel size 49-68.4 mu m). The radiation doses of the PCD-CT scans were varied from 2.5 to 120 mGy based on the volume CT dose index (CTDIvol32). For the PCD-CT scans, contrast-to-noise ratio and trabecular sharpness were calculated and compared between radiation doses. mu CT and PCD-CT scans were registered. The trabecular bone was then segmented from all PCD-CT and mu CT scans and split into cubes with 6mm edge length. For each cube, bone volume over total volume, trabecular thickness, trabecular number, and trabecular heterogeneity were calculated and compared between corresponding PCD-CT and mu CT cubes. Results: With increasing dose, contrast-to-noise ratio and trabecular sharpness values increased for the PCD-CT images. Already at the lowest dose, high correlations between the trabecular microarchitectural parameters between mu CT and PCD-CT were found (R2 = 0.55-0.95), which improved with increasing radiation dose (R2 = 0.76-0.96 at 20 mGy). Conclusions: PCD-CT can be used to quantify trabecular bone microarchitecture, with accuracy comparable to mu CT and at clinically relevant radiation doses.