The Assisted Heart: Function and flow during left ventricular assistance
Linköping University, Department of Health, Medicine and Caring Sciences, Division of Diagnostics and Specialist Medicine. Linköping University, Faculty of Medicine and Health Sciences. Linköping University, Center for Medical Image Science and Visualization (CMIV). Region Östergötland, Heart Center, Department of Thoracic and Vascular Surgery.ORCID iD: 0000-0002-8581-1050
2025 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Durable left ventricular assist device (LVAD) therapy has been revolutionised by the HeartMate 3™ (HM3), a fully magnetically levitated centrifugal pump, offering superior haemocompatibility and durability. However, long-term outcomes remain limited by complications such as aortic regurgitation (AR), right ventricular failure (RHF), and extrinsic outflow graft obstruction (eOGO). The current clinical standard for haemodynamic monitoring, echocardiography, is restricted by poor acoustic windows caused by the metallic body of the device. The overarching aim of this thesis was to investigate how HM3 therapy interacts with the native heart´s function and haemodynamics, employing photon-counting computed tomography (PCCT) as the foundation for new advanced imaging approaches.

The thesis comprises five papers. Paper I systematically reviewed echocardiographic evidence in HM3 recipients specifically, identifying AR, RHF, and interventricular interplay as key unresolved challenges. Paper II optimised PCCT acquisition and reconstruction protocols, demonstrating that tailored strategies can markedly reduce metal artefacts. Paper III combined PCCT-derived geometries with computational fluid dynamics (CFD) in cases of suspected eOGO, showing that anatomical narrowing does not invariably result in haemodynamic compromise. Paper IV applied a more advanced CFD-modelling approach at different degrees of AR, revealing that even mild regurgitation can induce a blind-loop circulation, a haemodynamic phenomenon with substantial loss of effective systemic output when the loop-circulation increases. Paper V analysed right ventricular volumetric variables across varying pump speeds, demonstrating progressive dilatation and reduced ejection fraction at higher speeds, with impaired contractile reserve – changes that could not be detected by concomitant echocardiographic variables.

In conclusion, by integrating artefact-reduced PCCT with image processing and flow modelling, this thesis provides mechanistic insights into device-heart interactions that cannot be captured by conventional methods. The findings indicate that radiologically identified graft stenoses may lack haemodynamic significance, that AR can markedly reduce effective output through blind-loop circulation, and that RV maladaptation at higher pump speeds may remain unnoticed by standard monitoring.  

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2025. , p. 73
Series
Linköping University Medical Dissertations, ISSN 0345-0082 ; 1999
National Category
Cardiology and Cardiovascular Disease
Identifiers
URN: urn:nbn:se:liu:diva-218257DOI: 10.3384/9789181182545ISBN: 9789181182538 (print)ISBN: 9789181182545 (electronic)OAI: oai:DiVA.org:liu-218257DiVA, id: diva2:2002811
Public defence
2025-11-06, Granitsalen, building 440, Campus US, Linköping, 13:00 (English)
Opponent
Supervisors
Note

Funding: This work has been conducted in collaboration with the Center for Medical Image Science and Visualization (CMIV) at Linköping University, Sweden. CMIV is acknowledged for the provision of financial support and research infrastructure.

The computations were enabled by resources provided by the National Academic Infrastructure for Supercomputing in Sweden (NAISS), the Swedish National Infrastructure for Computing (SNIC), and the National Supercomputer Centre (NSC). 

Furthermore, the Foundation Linköpings Läkarsällskap and Region Östergötland are acknowledged for financial support.

Available from: 2025-10-02 Created: 2025-10-02 Last updated: 2025-10-02Bibliographically approved
List of papers
1. Echocardiographic haemodynamic monitoring in the context of HeartMate 3™ therapy: a systematic review
Open this publication in new window or tab >>Echocardiographic haemodynamic monitoring in the context of HeartMate 3™ therapy: a systematic review
Show others...
2024 (English)In: ESC Heart Failure, E-ISSN 2055-5822, Vol. 11, no 4, p. 2033-2042Article in journal (Refereed) Published
Abstract [en]

AimsWhile echocardiography remains essential within haemodynamic monitoring of durable mechanical circulatory support, previous echocardiographic guidelines are missing scientific evidence for the novel HeartMate 3 (TM) (HM3) system. Accordingly, this review aims to summarize available echocardiographic evidence including HM3.Methods and resultsThis systematic review adhered to the PRISMA 2020 guidelines. Searches were conducted during August 2023 across PubMed, Embase, and Google Scholar using specific echocardiographic terms combined with system identifiers. Study quality was assessed using the Newcastle-Ottawa Scale (NOS) for cohort studies and Critical Appraisal Instrument (PCAI) for cross-sectional studies. Nine studies met the inclusion criteria, of which eight cohort studies and one cross-sectional study. Aortic regurgitation (AR) prevalence at approximately 12 months of support exhibited heterogenicity (33.5% (Delta 33%)) in a limited number of studies (n = 3). Several studies (n = 5) demonstrated an increasing prevalence and severity of AR during HM3 support, generating moderate to high level of evidence. One AR study showed a higher cumulative incidence of death and heart failure (HF) readmission compared with those without significant AR, hazard ratio 3.42 (95% CI 1.48-8.76). A second study showed that a worsening AR group had significantly lower survival-free from HF readmission (59% vs. 89%, P = 0.023) with a hazard ratio of 5.18 (95% CI 1.07-25.0), while a third study did not reveal any differences in cardiac-related hospitalizations in the 12 months follow-up or non-cardiac-related hospitalization. Mitral regurgitation (MR) prevalence at approximately 12 months of support exhibited good consistency 15.0% (Delta 0.8%) in both included studies, which did not reveal any significant pattern of changing prevalence over time. Tricuspid regurgitation (TR) prevalence at approximately 12 months of support exhibited fair consistency 28.5% (Delta 8.3%) in a limited number of studies (n = 2); both studies showed a statistically un-confirmed trend of increased TR prevalence over time. The evidence of general prevalence of right ventricular dysfunction (RVD) was insufficient due to lack of studies.ConclusionsThere are few methodologically consistent studies with focus on long-term haemodynamic effects. Aortic regurgitation still seems to be a prevalent and potentially significant finding. The available evidence concerning right heart function is limited despite clinical relevance and potential prognostic value. Potential interventricular and haemodynamic interplay are identified as a white field for future research.

Place, publisher, year, edition, pages
WILEY PERIODICALS, INC, 2024
Keywords
Echocardiography; Haemodynamic monitoring; Heart failure; HeartMate 3; LVAD; MCS
National Category
Cardiology and Cardiovascular Disease
Identifiers
urn:nbn:se:liu:diva-202251 (URN)10.1002/ehf2.14759 (DOI)001189296600001 ()38520314 (PubMedID)2-s2.0-85189200653 (Scopus ID)
Note

Funding Agencies|Region stergtland

Available from: 2024-04-09 Created: 2024-04-09 Last updated: 2025-10-02
2. Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in-depth assessment of metal artifacts
Open this publication in new window or tab >>Optimization of photon counting CT for cardiac imaging in patients with left ventricular assist devices: An in-depth assessment of metal artifacts
Show others...
2024 (English)In: Journal of Applied Clinical Medical Physics, E-ISSN 1526-9914, Vol. 25, no 7, article id e14386Article in journal (Refereed) Published
Abstract [en]

PurposePhoton counting CT (PCCT) holds promise for mitigating metal artifacts and can produce virtual mono-energetic images (VMI), while maintaining temporal resolution, making it a valuable tool for characterizing the heart. This study aimed to evaluate and optimize PCCT for cardiac imaging in patients during left ventricular assistance device (LVAD) therapy by conducting an in-depth objective assessment of metal artifacts and visual grading.MethodsVarious scan and reconstruction settings were tested on a phantom and further evaluated on a patient acquisition to identify the optimal protocol settings. The phantom comprised an empty thoracic cavity, supplemented with heart and lungs from a cadaveric lamb. The heart was implanted with an LVAD (HeartMate 3) and iodine contrast. Scans were performed on a PCCT (NAEOTOM Alpha, Siemens Healthcare). Metal artifacts were assessed by three objective methods: Hounsfield units (HU)/SD measurements (DiffHU and SDARTIFACT), Fourier analysis (AmplitudeLowFreq), and depicted LVAD volume in the images (BloomVol). Radiologists graded metal artifacts and the diagnostic interpretability in the LVAD lumen, cardiac tissue, lung tissue, and spinal cord using a 5-point rating scale. Regression and correlation analysis were conducted to determine the assessment method most closely associated with acquisition and reconstruction parameters, as well as the objective method demonstrating the highest correlation with visual grading.ResultsDue to blooming artifacts, the LVAD volume fluctuated between 27.0 and 92.7 cm3. This variance was primarily influenced by kVp, kernel, keV, and iMAR (R2 = 0.989). Radiologists favored pacemaker iMAR, 3 mm slice thickness, and T3D keV and kernel Bv56f for minimal metal artifacts in cardiac tissue assessment, and 110 keV and Qr40f for lung tissue interpretation. The model adequacy for DiffHU SDARTIFACT, AmplitueLowFreq, and BloomVol was 0.28, 0.76, 0.29, and 0.99 respectively for phantom data, and 0.95, 0.98, 1.00, and 0.99 for in-vivo data. For in-vivo data, the correlation between visual grading (VGSUM) and DiffHU SDARTIFACT, AmplitueLowFreq, and BloomVol was -0.16, -0.01, -0.48, and -0.40 respectively.ConclusionWe found that optimal scan settings for LVAD imaging involved using 120 kVp and IQ level 80. Employing T3D with pacemaker iMAR, the sharpest allowed vascular kernel (Bv56f), and VMI at 110 keV with kernel Qr40 yields images suitable for cardiac imaging during LVAD-therapy. Volumetric measurements of the LVAD for determination of the extent of blooming artifacts was shown to be the best objective method to assess metal artifacts.

Place, publisher, year, edition, pages
WILEY, 2024
Keywords
heart; left ventricular assistance device; metal artifact; myocardium; photon-counting computed tomography
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
urn:nbn:se:liu:diva-203749 (URN)10.1002/acm2.14386 (DOI)001219658800001 ()38739330 (PubMedID)2-s2.0-85192926572 (Scopus ID)
Note

Funding Agencies|CMIV; Swedish Research Council [2022-03931]; Swedish Heart and Lung Foundation [20210441]; ALF Grants Region Ostergotland [ROE-987498]; Sweden's Innovation Agency VINNOVA [2019-02261]; Foundation Linkoepings Laekaresaellskap Research Fund

Available from: 2024-05-28 Created: 2024-05-28 Last updated: 2026-03-13
3. Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.
Open this publication in new window or tab >>Haemodynamic significance of extrinsic outflow graft stenoses during HeartMate 3™ therapy.
Show others...
2024 (English)In: European heart journal. Imaging methods and practice, ISSN 2755-9637, Vol. 2, no 3, article id qyae082Article in journal (Refereed) Published
Abstract [en]

AIMS: The HeartMate 3 (HM3) implantable left ventricular assist device connects the left ventricle apex to the aorta via an outflow graft. Extrinsic obstruction of the graft (eOGO) is associated with serious morbidity and mortality and recently led to a Food and Drug Administration Class 1 device recall of HM3. This study aimed to provide a better understanding of the haemodynamic impact of extrinsic stenoses.

METHODS AND RESULTS: Computed tomography (CT) images of two retrospectively identified patients with eOGO (29 and 36% decrease in cross-sectional area, respectively, by radiological evaluation) were acquired with a novel photon-counting CT system. Numerical evaluations of haemodynamics were conducted using a high-fidelity 3D computational fluid dynamics approach on both the patient-specific graft geometries and in two virtually augmented stenotic severities and three device flows. Visual analysis identified increased velocity, pressure, and turbulent flow in the outer anterior curvature of the outflow graft; however, changes in graft pressure gradients were slight (1-9 mmHg) across the range of stenosis severities and flow rates tested.

CONCLUSION: Evidence of eOGO during HM3 support and the recent device recall can provoke clinical apprehension and interventions. The haemodynamic impact of a stenosis detected visually or by quantification of cross-sectional area reduction may be difficult to predict and easily overestimated. This numerical study suggests that, for clinically encountered flow rates and stenosis severities below 61% in cross-sectional area decrease, eOGO may have low haemodynamic impact. This suggests that patients without symptoms or signs consistent with haemodynamically significant obstruction might be managed expectantly.

Place, publisher, year, edition, pages
Oxford University Press, 2024
Keywords
CFD, HeartMate 3, LVAD, haemodynamics, photon-counting CT
National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-208765 (URN)10.1093/ehjimp/qyae082 (DOI)39224624 (PubMedID)
Available from: 2024-10-23 Created: 2024-10-23 Last updated: 2025-10-02Bibliographically approved

Open Access in DiVA

fulltext(5373 kB)302 downloads
File information
File name FULLTEXT01.pdfFile size 5373 kBChecksum SHA-512
4fe9cce05323e5f77edfa2d9033229e549ae81abdbc40d6dea38f59a591aca84770c347727a0736a56951c3b711ce91e54ef12e651c6d12dcb373bb3fd621505
Type fulltextMimetype application/pdf
Order online >>

Other links

Publisher's full text

Authority records

Ohlsson, Linus

Search in DiVA

By author/editor
Ohlsson, Linus
By organisation
Division of Diagnostics and Specialist MedicineFaculty of Medicine and Health SciencesCenter for Medical Image Science and Visualization (CMIV)Department of Thoracic and Vascular Surgery
Cardiology and Cardiovascular Disease

Search outside of DiVA

GoogleGoogle Scholar
Total: 307 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
isbn
urn-nbn

Altmetric score

doi
isbn
urn-nbn
Total: 3135 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf