A numerical framework for preprocedural prosthetic valve positioning and hemodynamic evaluationShow others and affiliations
2025 (English)In: Biomechanics and Modeling in Mechanobiology, ISSN 1617-7959, E-ISSN 1617-7940, Vol. 25, no 1, article id 3Article in journal (Refereed) Published
Abstract [en]
Aortic valve replacement is a cornerstone treatment for severe aortic valve diseases, including stenosis and regurgitation. Suboptimal valve seating can elevate the transvalvular pressure gradient, while valve orientation and size may produce flow jets that impinge on the ascending aorta, potentially weakening the vessel wall. Such hemodynamic complications can compromise valve performance and patient outcomes. This study presents a computational fluid dynamics framework, derived from medical CT images, for preprocedural hemodynamic assessment of aortic valve replacement. The framework minimizes user input and delivers rapid results, enabling efficient evaluation of valve types, orientations, and their hemodynamic impact. The results demonstrate that non-optimal implantation angles substantially increase pressure drop across the valve, thereby imposing higher workload on the heart. This automated and efficient simulation framework demonstrates strong potential for clinical application, supporting precise planning and execution of valve implantation procedures to improve patient care.
Place, publisher, year, edition, pages
Springer Nature , 2025. Vol. 25, no 1, article id 3
Keywords [en]
Aortic valve replacement; Computational fluid dynamics; Hemodynamics; Valve implantation planning; Cardiac CT; Medical image-based modeling
National Category
Fluid Mechanics
Identifiers
URN: urn:nbn:se:liu:diva-220128DOI: 10.1007/s10237-025-02025-7ISI: 001639408500001PubMedID: 41384982Scopus ID: 2-s2.0-105024656750OAI: oai:DiVA.org:liu-220128DiVA, id: diva2:2022030
Funder
Karolinska InstituteNational Supercomputer Centre (NSC), Sweden
Note
Funding Agencies|Linkping University; Karolinska Institute
2025-12-162025-12-162026-01-22