The MR quality landscape in Europe
Firenze Univ Hosp, Italy.
Univ Pavia, Italy; IRCCS Mondino Fdn, Italy.
IRCCS, Italy.
Univ Coll Cork, Ireland; European Federat Radiographer Soc EFRS, Portugal.
Show others and affiliations
2026 (English)In: Insights into Imaging, E-ISSN 1869-4101, Vol. 17, no 1, article id 125Article in journal (Refereed) Published
Abstract [en]

Objectives Despite the absence of ionizing radiation, magnetic resonance (MR) has inherent risks in clinical practice that can have serious health consequences if overlooked. At an international level, there are MR safety guidelines that help define the organization of a radiology department to minimize the risks for patients and personnel. However, competing guidelines exist and not every country and institution adheres to the same standards. In this work, we aim to understand the current situation regarding MR safety practices across Europe, and to identify the points where harmonization, coordination, or further education is needed. Methods An anonymous survey questionnaire was distributed between April and June 2023 through ESR member societies to healthcare professionals, aimed to assess personnel training, local policies, scanning practices, and accidents. Results Seven hundred and ninety-three responses were obtained from 44 different countries. The majority of respondents from five countries reported that MR safety is mandated by law, but we could only confirm two (Italy and Austria). While 77% of the responses said that their institution had a clear MR safety guideline, 52% said that nobody in their institution had received specific MR safety training. MR-conditional cardiac devices are mostly scanned in university hospitals (reported by 75% of respondents from this type of institution) but in only 42% of outpatient facilities. MR-unsafe cardiac devices are only scanned off-label in 27% of university hospitals, and in an even smaller share of other institutions. Approximately 12% of the respondents reported MR-related accidents resulting in patient or personnel injury. Overall, there is the sentiment that MR safety education and regulation are needed. Conclusions The European landscape in terms of MR safety is very heterogeneous, with different regulations across countries, and different procedures for MR safety training and their application in clinical routine. The European Society of Radiology is optimally positioned to play an active role in the harmonization of MR safety education and practices across Europe, and we are proposing a four-tiered framework for the development of a teaching curriculum for MR safety training.

Place, publisher, year, edition, pages
SPRINGER WIEN , 2026. Vol. 17, no 1, article id 125
Keywords [en]
Magnetic resonance imaging; Quality assurance, health care; Quality control; Education; Guideline adherence
National Category
Nursing
Identifiers
URN: urn:nbn:se:liu:diva-224063DOI: 10.1186/s13244-026-02280-xISI: 001759028400001PubMedID: 42096163OAI: oai:DiVA.org:liu-224063DiVA, id: diva2:2060610
Available from: 2026-05-18 Created: 2026-05-18 Last updated: 2026-07-01

Open Access in DiVA

fulltext(1210 kB)8 downloads
File information
File name FULLTEXT01.pdfFile size 1210 kBChecksum SHA-512
fac5933d00937dc601c52a69fe8b17372230015e2eea60888fc7cce05ccd8aecc9c46e5f50fbf4282d07ceacde56ebc187e06a99a04eb7558996a33580ca2c3a
Type fulltextMimetype application/pdf

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Lundberg, Peter
By organisation
Division of Diagnostics and Specialist MedicineFaculty of Medicine and Health SciencesMedical radiation physicsDepartment of Radiology in LinköpingCenter for Medical Image Science and Visualization (CMIV)
In the same journal
Insights into Imaging
Nursing

Search outside of DiVA

GoogleGoogle Scholar
Total: 8 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 22 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf