Introduction With the failure of high-density lipoprotein (HDL) cholesterol raising therapies to reduce cardiovascular risk, attention has turned towards HDL composition and function. There are several techniques for the isolation of lipoproteins from plasma, with density ultracentrifugation considered the gold-standard method. It is not known which form of density ultracentrifugation is optimal for HDL composition and function analyses.Methods This study compared HDL composition by nLC-MS/MS and ELISA, subclass distribution by gel electrophoresis, and in vitro vascular anti-inflammatory function in HDL fractions isolated by sodium bromide sequential density ultracentrifugation (SDU) and iodixanol density gradient ultracentrifugation (DGU).Results HDL composition differed between the two isolation techniques, with DGU-isolated HDL fractions containing a higher total protein content than SDU (10.14 +/- 1.49 mg/mL compared to 3.18 +/- 1.10 mg/mL respectively, p < 0.001, mean +/- SD) but lower proteomic detection of key HDL proteins such as apolipoprotein A-I, apolipoprotein A-II and paraoxonase-1 in DGU-isolated HDL. HDL subclass distribution could not be determined in DGU HDL fractions due to contaminating plasma proteins. Vascular anti-inflammatory function was higher in DGU HDL (77.5 +/- 4.4 % compared to 58.3 +/- 4.8%, mean +/- SD, p = 0.014) possibly due to the presence of confounding plasma proteins.Conclusion Further steps may need to be added to DGU methods to obtain a 'cleaner' HDL fraction and therefore currently SDU is better suited to studies of HDL composition and function.
Funding Agencies|British Heart Foundation