A simple model for estimating the particle size dependence of absolute efficiency of fluorescent screens
Radiation Physics Department, Borås Hospital, Borås.
Linköping University, Department of Medical and Health Sciences, Radiation Physics. Linköping University, Faculty of Health Sciences. Östergötlands Läns Landsting, Centre for Surgery, Orthopaedics and Cancer Treatment, Department of Radiation Physics UHL.ORCID iD: 0000-0003-0209-498X
1999 (English)In: Physics in Medicine and Biology, ISSN 0031-9155, E-ISSN 1361-6560, Vol. 44, no 5, p. 1353-67Article in journal (Refereed) Published
Abstract [en]

The absolute efficiency of a phosphor screen is the ratio of the light energy per unit area at the screen surface to the incident x-ray energy fluence. Particle size is a critical factor in determining the absolute efficiency, but in most models its influence is not accounted for. To allow derivation of the particle size dependence, a model is proposed that describes the optical properties of the screen by means of a single parameter, the light extinction factor, xi, and assumes that the intrinsic efficiency (light energy/energy imparted to the phosphor material) is independent of particle size. The value of xi depends on the type of screen (phosphor, reflective backing, coating and binder) and has to be determined from measurements on at least two screens with known particle size and thickness. The absolute efficiency can then be calculated for an extended range of particle sizes and/or screen thicknesses. To test the model, experimental data from the literature were used to derive values of xi for screens of La2O2S:Tb, LaOBr:Tm and ZnCdS:Ag. The extinction factor was found to vary between -6 and +20%. The non-physical negative value for xi, found from one set of experiments on La2O2S:Tb screens, may be explained as resulting from a lack of accurate knowledge of the actual tube potential, influencing calculated values of the energy imparted to the screen. The results are promising but further well-controlled experiments (including improved dosimetric calculations to account forescape of K-radiation from the screen) are needed to confirm the model.

Place, publisher, year, edition, pages
1999. Vol. 44, no 5, p. 1353-67
National Category
Medical and Health Sciences
Identifiers
URN: urn:nbn:se:liu:diva-70537DOI: doi:10.1088/0031-9155/44/5/319PubMedID: 10368024OAI: oai:DiVA.org:liu-70537DiVA, id: diva2:440151
Available from: 2011-09-12 Created: 2011-09-12 Last updated: 2021-03-24Bibliographically approved
In thesis
1. Radioluminescence: A simple model for fluorescent layers
Open this publication in new window or tab >>Radioluminescence: A simple model for fluorescent layers
2011 (English)Licentiate thesis, comprehensive summary (Other academic)
Abstract [en]

The aim of this thesis is to present a simple model for the radiation to light conversion processes in fluorescent layers as an aid in future developments and applications. Optimisation between sensitivity and spatial resolution for fluorescent layers in digital radiology is a delicate task where the extrinsic efficiency for various phosphors needs to be established for varying parameters. The extrinsic efficiency of a fluorescent layer can be expressed as the ratio of the light energy per unit area at the screen surface to the incident xray energy fluence. Particle size is a critical factor in determining the value of the extrinsic efficiency, but in most models it is not treated as an independent variable. Based on the definition of a light extinction factor (ξ), a model is proposed such that, knowing the intrinsic efficiency η, the particle size and the thickness of a certain make of screen, the extrinsic efficiency can be calculated for an extended range of particle sizes and / or screen thicknesses. The light extinction factor ξ is an optical parameter determined from experimental data on extrinsic efficiency. The proposed model is compared to established methods. Further experiments have confirmed the validity of the model. Monte-Carlo simulations have been utilised to refine the calculations of energy imparted to the phosphor by taking into account the escape of scattered and K-radiation generated in the screen and interface effects at the surfaces. The luminance was measured for a set of in-house manufactured screens of varying thicknesses and particle sizes. Utilising the proposed model, the corresponding calculated values deviated ± 14 %.within the studied range.

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2011. p. 36
Series
Linköping Studies in Health Sciences. Thesis, ISSN 1100-6013 ; 117
Keywords
Extrinsic efficiency, extinction factor, radioluminescence, modelling, Monte Carlo, optical attenuation, optimisation, particle size, phosphor, scintillator
National Category
Medical and Health Sciences
Identifiers
urn:nbn:se:liu:diva-70539 (URN)978-91-7393-098-7 (ISBN)
Supervisors
Available from: 2011-09-12 Created: 2011-09-12 Last updated: 2019-10-12Bibliographically approved
2. Radioluminescence: A simple model for fluorescent layers - analysis and applications
Open this publication in new window or tab >>Radioluminescence: A simple model for fluorescent layers - analysis and applications
2021 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

A phosphor or scintillator is a material that will emit visible light when struck by ionising radiation. In the early days of diagnostic radiology, it was discovered that the radiation dose needed to get an image on a film, could be greatly reduced by inserting a fluorescent layer of a phosphor in direct contact with the film. Thus, introducing the step of converting the ionising radiation to light in a first step. Going forward in time, film has been replaced with photodetectors and there is now a variety of imaging x-ray systems, still based on phosphors and scintillators. 

There is continuous research going on to optimise between the radiation dose needed and a sufficient image quality. These factors tend to be in opposition to each other. It is a complicated task to optimise these imaging system and new phosphor materials emerges regularly. One of the key factors is the efficiency of the conversion from xrays to light. In this work this is denoted “extrinsic efficiency”. It is important since it largely determines the final dose to the patient needed for the imaging task. Most imaging x-ray detectors are based on phosphor or scintillator types where their imaging performance has been improved through tweaking of various parameters (light guide structure, higher density, light emission spectrum matching to photodetectors, delayed fluorescence quenching etc) 

One key factor that largely determines the extrinsic efficiency of a specific phosphor is the particle size. Larger particles result in a higher luminance of the phosphor for the same radiation dose as does as a thicker phosphor layer (to a limit). There exists already a battery of models describing various phosphor qualities. However, particle size and thickness have not been treated as a fully independent variables in previous model works. Indirectly, the influence of these parameters is accounted for, but the existing models were either considered too general, containing several complex parameters and factors to cover all kind of cases or too highly specialised to be easily applicable to fluorescent detectors in diagnostic radiology. 

The aim of this thesis is therefore to describe and assess a simple model denoted the “LAC-model” (after the original authors Lindström and Alm Carlsson), developed for a fluorescent layer using individual sub-layers defined by the particle size diameter. The model is thought to be a tool for quickly evaluating various particle size and fluorescent layer thickness combinations for a chosen phosphor and design. It may also serve as a more intuitive description of the underlying parameters influencing the final extrinsic efficiency. 

Further tests affirmed the validity of the model through measurements. The LACmodel produced results deviating a maximum of +5 % from luminescence measurements. 

During the development of the model various assumptions and simplifications were made. One assumption was the absence of a so called “dead layer”. This is a layer supposedly surrounding each particle decreasing the efficiency of converting x-rays to light. It is not completely “dead” as in inactive but is thought to have a reduced efficiency. This phenomenon was struggled with, when historically designing electron beam stimulated phosphors for various applications (i.e. displays, TV tubes etc). There are also articles reporting dead layer influence for x-ray detectors (usually spectrometers i.e. not for imaging). By introducing a dead layer in the LAC-model the effect of the layer was investigated and was found to result in a change of less than 8% for the extrinsic efficiency. 

It was also noted that sometimes a dead layer effect may emerge at surfaces of a scintillator slab but not necessarily connected to the phosphor particles themselves. Due to differences between phosphor material and the surroundings, an interface effect arose to compete with the process of inherent dead layers of the individual particles. It was found to be mostly negligible for x-rays in the studied energy and material range. However, an effect was shown for electrons as incident ionising radiation which could shed some light on the strangely neglected apparent dead layer created this way. Finally, applications, one involving developing a prototype for checking the light field radiation field coincidence, were evaluated for overall performance and the optimisation level of the applied fluorescent layer. Interesting findings were made during the development process: for the first time to the knowledge of the author, focus shift wandering was quantified in the corresponding movement of the x-ray field edge and a non-trivial discussion on the concept of an apparent light field edge resulted in a modified definition of the same.  

Abstract [sv]

En fosfor eller scintillator är ett material som avger synligt ljus när det träffas av joniserande strålning. Inom diagnostisk radiologi upptäckte man i ett tidigt skede att stråldosen som behövdes för att få en bild på en röntgenfilm, reducerades kraftigt om man placerade ett fluorescerande skikt, en fosfor, i direkt kontakt med filmen. I nutid har film ersatts med fotodetektorer och det finns nu en mängd olika röntgenbildsystem men som fortfarande är baserade på fosforer och scintillatorer. 

Det pågår en kontinuerlig forskning för att optimera mellan erforderlig stråldos och en tillräcklig god diagnostisk bildkvalitet. Dessa faktorer tenderar att motverka varandra. Det är en komplicerad uppgift att optimera röntgenbildsystemen och nya fosformaterial dyker ständigt upp. En av de viktiga egenskaperna är fosforns omvandlingseffektivitet från röntgen till ljus. I detta arbete används benämningen ”extrinsisk (yttre) effektivitet". Denna egenskap är viktig eftersom den i stor utsträckning bestämmer den slutliga dosen till patienten som krävs för bilddiagnostiken. De flesta röntgendetektorer är baserade på fosfor- eller scintillatortyper där bildprestanda har förbättrats genom att utveckla olika parametrar (ljusledarstruktur, högre densitet, ljusemissionsspektrum som matchar fotodetektorer, minskad efterlysning etc.). En viktig faktor som i stor utsträckning bestämmer omvandlingseffektiviteten hos en specifik fosfor är partikelstorleken. Större partiklar resulterar i en högre luminescens (mer ljus) från fosforen för samma stråldos. Vilket också gäller för ett tjockare fosforlager (till en viss gräns!). Det finns redan fysikaliska modeller som beskriver olika fosforparametrar men partikelstorlek och fosfortjocklek har dock inte hanterats som fristående variabler i dessa modellarbeten. Istället har deras inverkan modellerats indirekt men det har gjort att de befintliga modellerna kan anses komplexa. De är antingen för generella som medför flera komplexa parametrar och faktorer för att täcka alla tänkbara varianter eller för specialiserade för att kunna tillämpas enkelt på fluorescerande detektorer i diagnostisk radiologi. 

Syftet med denna avhandling är därför att beskriva och analysera en praktisk modell betecknad ”LAC-modellen” (efter de ursprungliga författarna Lindström och Alm Carlsson). Den är utvecklad för ett fluorescerande block som består av flera underliggande skikt vars tjocklek bestäms av partiklarnas diameter. Avsikten med modellen är att den ska vara ett verktyg för att snabbt utvärdera olika varianter av partikelstorlek och tjockleks-kombinationer för en vald fosfor med i grunden samma design. Experiment har bekräftat modellens giltighet och mätresultat visar att modellresultaten avvek maximalt +5% från luminiscensmätningar. 

Utvecklingen av modellen krävde olika antaganden och förenklingar. Ett antagande var frånvaron av ett så kallat ”dött lager”. Det är ett skikt som antas omge varje partikel och som därför minskar omvandlingseffektiviteten från röntgen till ljus. Det är dock inte helt "dött" i meningen helt inaktivt men har en mindre förmåga att omvandla röntgen till ljus jämfört med fosforns huvudmaterial. Historisk sett har man försökt åtgärda detta fenomen under lång tid och speciellt för applikationer där man använt sig av elektronstrålar (dvs olika typer av displayer, TV-rör etc.). Just för elektroner har man sett att döda skiktet tenderar att växa med tiden. Det finns också artiklar som rapporterar en påverkan av röntgendetektorers funktion (vanligtvis dock för spektrometrar, dvs inte för avbildning). 

Genom att införa ett dött skikt i LAC-modellen undersöktes skiktets effekt och visade sig resultera i en förändring på mindre än 8% för effektiviteten. Det noterades också att ibland kan en dödskiktsliknande effekt uppstå vid ytor av ett scintillatorblock men inte nödvändigtvis pga. av själva fosforpartiklarnas ljusomvandlingsegenskaper. Då det uppstår skillnader mellan fosformaterialet och omgivningen får man en s.k. gränsskiktseffekt som s.a.s. konkurrerar med kemiskt döda skiktet på de enskilda partiklarna. 

De döda skiktens inverkan visade sig i princip försumbara för röntgenbild-detektorer - åtminstone inom det studerade energi- och materialområdet. En tydlig effekt kunde dock noteras för joniserande strålning i form av elektroner. Simuleringarna kunde ge en bättre bild av egenskaperna hos det döda skiktet som skapats på detta sätt. 

Slutligen utvärderades två applikationer med hjälp av LAC-modellen: en prototyp för kontroll av ljusfältets och strålfältets överenstämmelse i läge och position. Samt en etablerad produkt med samma användningsområde. I båda fallen undersöktes det fluorescerande skiktets optimeringsgrad. Intressanta resultat noterades under utvecklingsprocessen av prototypen: för första gången, så vitt författaren vet, kunde man kvantifiera röntgenrörs s.k. fokusvandring.  

Place, publisher, year, edition, pages
Linköping: Linköping University Electronic Press, 2021. p. 61
Series
Linköping University Medical Dissertations, ISSN 0345-0082 ; 1773
National Category
Physical Sciences Radiology, Nuclear Medicine and Medical Imaging
Identifiers
urn:nbn:se:liu:diva-174573 (URN)10.3384/diss.diva-174573 (DOI)9789179296841 (ISBN)
Public defence
2021-04-27, Belladonna, Building 511, Campus US, Linköping, 13:00 (English)
Opponent
Supervisors
Available from: 2021-03-24 Created: 2021-03-24 Last updated: 2021-10-13Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textPubMed

Authority records

Alm Carlsson, Gudrun

Search in DiVA

By author/editor
Alm Carlsson, Gudrun
By organisation
Radiation PhysicsFaculty of Health SciencesDepartment of Radiation Physics UHL
In the same journal
Physics in Medicine and Biology
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 260 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • oxford
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf