Determination of the Real Surface Area of a Screen-Printed Electrode by ChronocoulometryShow others and affiliations
2012 (English)In: International Journal of Electrochemical Science, E-ISSN 1452-3981, Vol. 7, no 7, p. 6214-6220
Article in journal (Refereed) Published
Abstract [en]
A simple and effective method applying chronocoulometry and cyclic voltammetry to the determination of the real surface area of a screen-printed electrode, for the first time, is reported. The method is based on the Anson plot and the standard addition method. This method could greatly facilitate determination of the real surface area of commercial screen-printed electrodes, which still remains underdeveloped despite the critical importance of quality control in this industry.
Place, publisher, year, edition, pages
ELECTROCHEMICAL SCIENCE GROUP, A SPOMENICE 7/12 , 19210 BOR, BELGRADE, VJ 12, SERBIA , 2012. Vol. 7, no 7, p. 6214-6220
Keywords [en]
Screen-Printed Electrodes, Real/Active Surface Area, Chronocoulometry, Anson Plot, Standard Addition Method
National Category
Engineering and Technology
Identifiers
URN: urn:nbn:se:liu:diva-79984ISI: 000306399700038OAI: oai:DiVA.org:liu-79984DiVA, id: diva2:544948
Note
Funding Agencies|Pelikan Technologies, Inc.||
2012-08-172012-08-172024-01-17